real-time view display at any time, tap the
Exit
button to return to the
Real-Time View
menu.
Enabling Emergency Purge Capability
The HMS-1655L may be configured to enable a convenient feature
that allows the fume hood to be put into a maximum face velocity
mode, or
Emergency Purge
mode, for emergency situations where
the exhaust valve is forced to its maximum open position. To enable
the
Emergency Purge
mode at the main display of the HMS-1655L,
set the configuration dipswitch S2 accordingly (see table on page 30).
Setting position 2 of S2 to the ON position enables the
Emergency
Purge
feature, and results in the ‘Press for
Emergency Purge
’ button
appearing on the as shown in Figure 10.
Triatek reserves the right to change product specifications without notice.
HMS-1655L
QUICK START GUIDE
Figure 10. Tapping
the Emergency Purge
button instantly puts the
fume hood in maximum
face velocity mode.
Triatek reserves the right to change product specifications without notice.