RA660 Operation Manual
8509790 Rev. 2
51
18/04/2018
A green bar indicates that octave band is below the defined limit, a red bar indicates
that octave band is above the defined limit.The gray line above the bars indicates the
limit for each band. The numbers above the gray line are the same limit values
displayed numerically.
The limit for each band is set in the
Setting>Room Monitor
screen. Also on this screen
is the Pause Testing If Over Limits switch. If this switch is on the audiometers will
automatically pause the test if the room noise exceeds the set limit and automatically
restart when the noise drops below the limit.
5.10 Exporting and Importing Data
The RA660 Hearcon Software can export or import its database to an XML or CSV
format file. This allows the RA660 Hearcon software to interface with third party hearing
consevation programs. To use this feature select the
Utilities
tab.