Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The RA660 is intended for use in the electromagnetic environment specified below.
The customer or the user of the RA660 is should assure that it is used in such an
environment.
Immunity Test
IEC 60601 Test
level
Compliance
Electromagnetic
Environment-Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
+6 kV contact
+8 kV air
+6 kV contact
+8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should
be greater than 30%.
Electrical fast
transient/burst
IEC61000-4-4
+2 kV for power
supply lines
+1 kV for
input/output lines
+2 kV for power
supply lines
+1 kV for
input/output lines
Mains power quality
should be that of a typical
commercial or residential
environment.
Surge
IEC 61000-4-5
+1 kV differential
mode
+2 kV common
mode
+1 kV differential
mode
+2 kV common
mode
Mains power quality
should be that of a typical
commercial or residential
environment.
Voltage dips,
short interruptions
and voltage
variations on
power supply
lines
IEC 61000-4-11
< 5 % UT
(>95 % dip in UT)
for 0.5 cycle
40 % UT
(60 % dip in UT) for
5 cycles
70% UT
(30% dip in UT) for
25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
< 5 % UT (>95 %
dip in UT)
for 0.5 cycle
40 % UT (60 % dip
in UT) for
5 cycles
70% UT (30 % dip
in UT) for
25 cycles
<5 % UT
Mains power quality
should be that of a typical
commercial or residential
environment. In case of a
power failure the RA660
will automatically shut
down within 10 s. If the
user of the RA660
requires continued
operation during power
mains interruptions, it is
recommended that the
RA660 be powered from
an uninterruptable power
supply or its battery.
Power frequency
(50/60 Hz)
IEC 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields should be
at levels characteristic of
a typical location in a
typical commercial or
residential environment.
Note:
UT is the A.C. mains voltage prior to application of the test level.
Summary of Contents for RA660
Page 1: ...RA660 Operation Manual RA660 PC based group audiometer...
Page 2: ...RA660 Operation Manual...
Page 41: ...RA660 Operation Manual 8509790 Rev 2 39 18 04 2018 RA660 Daily Listening Check Screens...
Page 56: ...RA660 Operation Manual 8509790 Rev 2 54 18 04 2018...
Page 70: ...RA660 Operation Manual 8509790 Rev 2 68 18 04 2018...
Page 71: ...RA660 Operation Manual 8509790 Rev 2 69 18 04 2018 7 2 Import Data Format...