RA660 Operation Manual
8509790 Rev. 2
55
18/04/2018
6 Technical Specifications
6.1 Technical Data
The touchTymp is an active, diagnostic medical product
according to the class IIa of the EU medical directive
93/42/EEC.
ITEM
SPECIFICATION
Standards:
ANSI S3.6-2010: Type 4
IEC 60645-1 : Type 4
ISO 389
ES 60601-1 class I protection, class B applied parts,
American Standards for Medical Electrical Equipment
IEC 60601-1 class I protection, class B applied parts,
International Standards for Medical Electrical
Equipment
IEC 60601-1-2, Electromagnetic compatibility
–
Requirements and tests
CSA C22.2 # 60601-1-08
Medical Device Directive (MDD) to comply with “EC
Directive” 93/42/EEC
Test Frequencies:
250 Hz
– 8000 Hz
Maximum Test Levels:
100 dB
HL
250 Hz
– 8000 Hz
90 dB
HL
IP 30 6 kHz & 8 kHz
Headset
AC with Earphone TDH-39P or DD45
Insert Earphones
RADIOEAR IP 30
Attenuator Range:
-10 to 100 dB
HL
in 5 dB
HL
Steps
Test Signal:
Pure tone
Modulation:
Pulse Tone 200 mS on 200 mS off
Subject Response:
Handheld response switch
Response Window:
1.8 s from presentation of tone
Communication:
Talk over
Data Connection:
USB
Stimulus Functions:
Tone Presenter
Bio-acoustic Simulator
Response at 65 dBHL +/- 10 dB at 250 to 8000 Hz
Room Monitor
Octave band measurements 125 to 8000 Hz
Warm-up Time:
Less than 10 min after power on
Mode of Operation:
Continuous
Environment Conditions:
Operation
+15 °C
… +35 °C / +59 °F … +95 °F
Humidity: 30 %
… 90 %, non-
condensing
Air pressure: 80 kPa ... 104 kPa
Summary of Contents for RA660
Page 1: ...RA660 Operation Manual RA660 PC based group audiometer...
Page 2: ...RA660 Operation Manual...
Page 41: ...RA660 Operation Manual 8509790 Rev 2 39 18 04 2018 RA660 Daily Listening Check Screens...
Page 56: ...RA660 Operation Manual 8509790 Rev 2 54 18 04 2018...
Page 70: ...RA660 Operation Manual 8509790 Rev 2 68 18 04 2018...
Page 71: ...RA660 Operation Manual 8509790 Rev 2 69 18 04 2018 7 2 Import Data Format...