OMNI-Test Laboratories, Inc.
Thermal Metering System Calibration
Y Factor
Previous Calibration Comparision
Manufacturer:
Apex
Date
1/17/2018
Acceptable
Deviation (5%)
Deviation
Model:
XC-60-EP
y Factor
0.981
0.04905
0.007
Serial Number:
0702004
Acceptance
Acceptable
OMNI Tracking No.:
OMNI-00372
Calibrated Orifice:
FALSE
Current Calibration
Acceptable y Deviation
0.020
Average Gas Meter y
Factor
Average
Orifice
Meter
dH@
Maximum y Deviation
0.008
TRUE
0.974
N/A
Acceptable dH@ Deviation
N/A
Calibration Date:
07/16/18
Maximum dH@ Deviation
N/A
N/A
Calibrated by:
B. Davis
Acceptance
Acceptable
Calibration Frequency:
6 months
Next Calibration Due:
1/16/2019
Instrument Range:
1.000
cfm
Standard Temp.:
68
oF
Reference Standard *
Standard Press.:
29.92
"Hg
Standard
Model
Standard Test Meter
Barometric Press., Pb:
30.08
"Hg
Calibrator
S/N
OMNI-00001
Signature/Date:
7/16/2018
Calib. Date
27-Oct-16
Calib. Value
0.9977
y factor (ref)
Calibration Parameters
Run 1
Run 2
Run 3
Reference Meter Pressure ("H2O), Pr
0.00
0.00
0.00
DGM Pressure ("H2O), Pd
3.00
1.30
0.80
Initial Reference Meter
540.8
547.1
627.2
Final Reference Meter
547.004
552.102
633.4
Initial DGM
0
0
0
Final DGM
6.244
5.121
6.353
Temp. Ref. Meter (°F), Tr
86.0
88.0
88.0
Temperature DGM (°F), Td
85.0
87.0
87.0
Time (min)
27.5
34.5
56.5
Net Volume Ref. Meter, Vr
6.204
5.002
6.200
Net Volume DGM, Vd
6.244
5.121
6.353
Gas Meter y Factor =
0.982
0.970
0.970
Gas Meter y Factor Deviation (from avg.)
0.008
0.004
0.004
Orifice dH@
N/A
N/A
N/A
Orifice dH@ Deviation (from avg.)
N/A
N/A
N/A
where:
1. Deviation = |Average value for all runs - current run value|
** 2. y = [Vr x (y factor (ref)) x (Pb + (Pr/ 13.6)) x (Td + 460) ]/ [Vd x (Pb + (Pd / 13.6)) x (Tr + 460)]
** 3. dH@ = 0.0317 x Pd / (Pb (Td + 460)) x [ (Tr + 460) x time) / Vr ]^2
* Reference calibration is traceable to NIST through NIST Test # 40674, Kimble ASTM E1272, or NIST traceable laboratory
** Equations come from EPA Method 5
The uncertainty of measurement is ±0.14 ft
3
/min. This is based on the reference standard having a TAR (Test Accuracy Ratio) of at least 4:1.
Yes
Control No. C-SSB-0005, Effective Date: 7/7/2011
108 of 158
Summary of Contents for Xtrordinair 44 Elite
Page 36: ...36 of 158 ...
Page 37: ...37 of 158 ...
Page 49: ...49 of 158 ...
Page 59: ...59 of 158 ...
Page 60: ...60 of 158 ...
Page 69: ...69 of 158 ...
Page 70: ...70 of 158 ...
Page 84: ...84 of 158 ...
Page 85: ...85 of 158 ...
Page 89: ...89 of 158 ...
Page 90: ...90 of 158 ...
Page 91: ...91 of 158 ...
Page 92: ...92 of 158 ...
Page 93: ...93 of 158 ...
Page 94: ...94 of 158 ...
Page 95: ...95 of 158 ...
Page 96: ...96 of 158 ...
Page 97: ...97 of 158 ...
Page 98: ...98 of 158 ...
Page 99: ...99 of 158 ...
Page 101: ...101 of 158 ...
Page 102: ...102 of 158 ...
Page 103: ...103 of 158 ...
Page 104: ...104 of 158 ...
Page 106: ...7 20 18 106 of 158 ...
Page 109: ...7 20 18 109 of 158 ...
Page 111: ...111 of 158 ...
Page 112: ...112 of 158 ...
Page 113: ...113 of 158 ...
Page 115: ...115 of 158 ...
Page 130: ...130 of 158 ...
Page 151: ...Safety Label 21 Travis Industries 10 24 2018 93508094 44 Elite Owner s 151 of 158 ...