background image

2

Keyboard Functions

lb

kg

ZERO

NET

GROSS

Toggles between lb and kg
units if this key is enabled in
the User Menu.

Sets indicator to display "0"
when in Gross mode, and
within zero band range.

Toggles between Gross and
Net weight display.

Zero

G

N

lb

kg

Motion

ZERO

NET

GROSS

TARE

PRINT

Tare

CAPACITY :

300x0.1 lb
150x0.05 kg

lb

CLASS

max

= 3000

n

kg

SRP SERIES FRONT PANEL

TARE

Used to zero a weight indication in Net mode.

PRINT

Sends "Print" data to printer if scale is stable and not in overload. Not active when "Continuous" option
is selected in User Menu.

Configuration

All Serial Communication parameters are
embedded in a user menu consisting of 5
separate menu selections, each with its own
sub-menu of choices. If your scale is con-
nected by RS-232 serial communication
lines to other equipment, or you wish to dis-
able the 

lb/kg

 key, you may have to modify

some or all of these parameters.

To modify these parameters, you must first
enter the USER MENU mode. Once there,
four of the front panel keys become direc-
tional navigators to move around in the
menus, and one key is used to save or SET
the selections. Complete directions are
found below.

To place the unit in USER MENU mode:

1. Turn the scale off.

 
2. While holding down the 

lb/kg

key, turn the scale back on.

3. When the display shows "

A1

",

the unit is in USER MENU
mode, and you can release the

lb/kg

 key. Shown at right are the

directional and SET key assign-
ments.

Zero

G

N

lb

kg

Motion

SET

Tare

CAPACITY :

300x0.1 lb
150x0.05 kg

CLASS

max

= 3000

n

Summary of Contents for SRP-300

Page 1: ...SRP 300 Digital Bench Scale Operation Manual Revision 1 2 August 24 2000...

Page 2: ...ell Technology Inc Contents subject to change without notice Transcell Technology Inc 975 Deerfield Parkway Buffalo Grove IL 60090 Tel 847 419 9180 Fax 847 419 1515 E mail transcell transcell net Web...

Page 3: ...communication parameters that are embed ded in the User Menu The User Menu is accessed through the front panel keys Complete directions in cluding a graphical road map are found in the Configuration s...

Page 4: ...separate menu selections each with its own sub menu of choices If your scale is con nected by RS 232 serial communication lines to other equipment or you wish to dis able the lb kg key you may have t...

Page 5: ...t and one stop bit 7O 7 data bits with odd parity bit and one stop bit 7E 7 data bits with even parity bit and one stop bit 7 7 data bits with no parity bit and two stop bits 8 7O 7E 7 A3 Mode of Seri...

Page 6: ...e NET GROSS key to save the value 6 If the calibration was successful the display will show ECAL momentarily then freeze Exit the Calibration mode and enter the Normal Operating Mode by turning off th...

Page 7: ...0 C to 40 C OPTIONAL SWIVEL POST 18 Tall Stainless Steel Construction POWER SOURCE AC Adapter 12VDC 800mA included WEIGHT Net Weight 45 lb 20 4 kg Shipping Weight 50 lb 22 7 kg Certificate of Conforma...

Page 8: ...the nonconformity actually exists and was not caused by accident misuse neglect alteration improper installation improper or unau thorized repair or improper testing and 4 Such goods have not been mod...