3
Configuration /
Continued
USER MENU CHART
A1
Baud Rate
1200
2400 4800 9600
A2
Data Bits, Parity
8
7o
7E
A3
Transmission Mode
C
d
A4
Serial Data Format
A5
Disable lb/kg Key
1
0
0
1
2
3
19200
7
4
To place the unit back into the NORMAL OPERATING mode, turn off the power to the
unit.
With no keys held down, turn the power back on. All front panel keys will now return
to their normal mode of operation.
User Menu Descriptions
NAME/CODE
DESCRIPTION
CODE/VALUE
A1
Baud Rate
Selects the baud rate for data transmission through the serial port.
1200
2400
√√
4800 9600
19200
A2
Data Bits and
Parity
Selects the number of data bits and parity of serial transmission.
"8" = 8 data bits with no parity bit and one stop bit
"7O" = 7 data bits with odd parity bit and one stop bit
"7E" = 7 data bits with even parity bit and one stop bit
"7" = 7 data bits with no parity bit and two stop bits
8
√√
7O
7E
7
A3
Mode of Serial
Transmission
Selects when data will be sent out of the serial port to a printer or
computer:
"C" = Continuous mode; send data continuously
"d" = Demand mode; send data when a PRINT command is issued
from the printer, computer, or indicator.
C
√√
d
A4
Serial Data
Format
Selects the data format to be transmitted via the serial port to a
printer or computer.
"0" = Consolidated Controls Format
"1" = Toledo 8213 Format "2" = NCI 3825Format
"3" = Transcell Technology Format
"4" = Fairbanks 70-2453-4 Format
0
1
2
3
√√
4
A5
Disable the
lb/kg Key
Allows the lb/kg key to be disabled so that an operator cannot acci-
dentally press the key and change the displayed units.
"0" = Disable the lb/kg key "1" = Enable the lb/kg key
0
1
√√