Trans instruments WalkLAB HP 9010 Operation Manual Download Page 9

 

8

 

7. 

CALIBRATION:

 

 
 
 
 
 
 
 
 

7.1

 

Prepare  standard  solutions  of  at  least  2  calibration  points.    If 
measurements are to be made in the acidic range, select 2 points 
between  pH-2  to  7.    If  measurements  are  made  in  the  alkaline 
range,  then  select  between  pH7  to  16.    If  full  range  is  required, 
calibrate at least 3 to 5 calibration points between pH-2 to 16. 

7.2 

Buffer standard must be set during

 

SETUP

 

in page 6, 

BuF

BuF

BuF

BuF

7.3

 

Standard  solutions  are  in  3  groups  of  standards  –

  ISO,  Nist 

and 

custom standards. 

 

ISO 

buffer standards are pH1.68, 7.00, 4.01, 10.01 and 12.45 

 

NIST 

buffer standards are pH1.68, 4.01, 6.86, 9.18 and 12.45 

 

Custom standards are defines manually for each buffer standard. 

7.4

 

Do  keep  in  mind  that  buffer  all  standards  are  specifies  at  25°C.    Calibration 
value should set to the exact value as the buffer’s solution value correlate to its 
current temperature. 

Calibration with ISO Standard 

7.5

 

Make sure the unit is properly installed and electrode connected. 

7.6

 

This  meter  can  automatically  recognize  the 

ISO

  Standard  buffer 

solution.  You will need at least 2 buffer solutions.  

Always begin with 

“Offset” calibration first

7.7

 

Rinse  the  pH  electrodes  and  temperature  probe  in  distilled-water 
then  dip  it  in  the  buffer  solution,  first  in  the  offset  buffer  (pH7.00) 
then subsequently each of the slope buffer solution. 

 

 

 

 

7.8

 

Press 

 CAL 

 key for 2 seconds and display will show the 

7.00

 

and 

 

CAL 

 

blinking, indicating it is in calibration mode. 

* Calibration should be performed as frequently as possible 

to ensure accurate measurement, depending on the 

frequency of tests performed.  Additional calibration 

solution should be purchased for future needs

.

 

 

Always rinses the electrode with distilled water before and after each 

test.  This is to prevent solution carry over or cross contamination.  

Standard solutions must maintain highest purity; otherwise the 

meter’s accuracy could be compromised. 

Summary of Contents for WalkLAB HP 9010

Page 1: ...0 15 0 11 4 1 14 5 10 13 2 9 6 3 12 7 8 15 0 15 0 1 14 1 14 13 2 13 2 3 12 3 12 11 4 11 4 5 10 5 10 9 6 9 6 7 8 7 8 WalkLAB Professional pH Meter HP 9010 Operations Manual...

Page 2: ...2 online data logging to PC sold separately 2 CONTENT Carefully unpack the box It should contain the following items a Main unit b Combination pH electrode c Temperature probe d Calibration buffer sol...

Page 3: ...to snap down cover 4 PRODUCT LAYOUT pH Electrode Socket Temperature Probe Socket Battery Cover Lift up latch here to open Battery Cover On Off Button Press once to switch ON To switch OFF press and h...

Page 4: ...and Meter mS ppt ppm kPa ION DO Cond Salinity TDS S ATC Auto Lock AM PM HOLD pH Button keys legend correlate to button key Indicator will blink while the function mode is active Date or Time display L...

Page 5: ...tomatically sense a stabilized reading and locks the endpoint reading Factory preset is Yes Yes Yes Yes 1 Press the ENTER key to set display will show Yes Yes Yes Yes blinking 2 Press the SELECT key u...

Page 6: ...last calibration information about the electrode conditions A good pH electrode should have an Offset value no greater than 60mV and a Slope value of above 75 and below 115 Otherwise the electrode sh...

Page 7: ...ELECT key up or down to choose between C C C C and F F F F 3 Press the ENTER key to confirm 4 Press the SELECT key up for next setting Avto off AUTO OFF TIME SETTING This setting allow you set the tim...

Page 8: ...ENTER clock is up to date 5 Press the SELECT key up for next setting RST MASTER RESET This setting allows you to reset the unit to the original factory s default Original factory default Buffer Stand...

Page 9: ...current temperature Calibration with ISO Standard 7 5 Make sure the unit is properly installed and electrode connected 7 6 This meter can automatically recognize the ISO Standard buffer solution You w...

Page 10: ...bration cannot proceed until they are rectified 7 11 When the meter beeps Offset point is established 7 12 Now 1 68 4 01 10 01 and 12 45 will alternate on main display 7 13 Rinse the pH electrodes and...

Page 11: ...uffer solution Temperature C Slope 1 68 Slope 4 01 Offset 6 86 Slope 9 18 Slope 12 45 0 1 67 4 01 6 98 9 47 13 43 5 1 67 4 00 6 95 9 38 13 21 10 1 67 4 00 6 92 9 32 13 00 15 1 67 4 00 6 90 9 27 12 81...

Page 12: ...press EXIT for 2 seconds will abort all Calibration with Custom Standard 7 38 Make sure the unit is properly installed and electrode connected 7 39 Rinse the pH electrodes and temperature probe in dis...

Page 13: ...t reading is established You can now record the reading 8 5 Press the READ button for each new measurement Measurement with Auto Lock feature disabled see Page 4 8 6 When feature is disabled meter wil...

Page 14: ...Depend on the liquid that is measured some volatile liquid may take up to 20 minutes or even longer time to stabilize Disable the Auto lock Page4 and Auto Off Page6 features Then endpoint reading shou...

Page 15: ...asure liquid within specified range ErLc Prob Readings drift or unstable cannot auto lock during Read or Cal 1 Ground loop in solution 2 Electrode is clog or expired 3 Electromagnetic interference 4 O...

Page 16: ...nsinstruments com http www transinstruments com Singapore Pte Ltd email sales transinstruments com Website www transinstrumnets com ISO9001 Certified Firm Rev 4 Sep2015 PRINTED IN SINGAPORE Quality ch...

Reviews: