Trans instruments WalkLAB HP 9010 Operation Manual Download Page 2

 

1

 

INTRODUCTION 

Your  purchase  of  this  professional  pH  meter  marks  a  step  forward  for 
you into the field of precision measurement.  Although this meter is a 
complex and delicate instrument, its usability will allow many years of 
use if proper operating techniques are observe and practice. 
 

1.

 

FEATURES: 

 

Multiple display of pH, mV and temperature 

 

Automatic buffer recognition with built in ISO and NIST standard 

 

Manual setting of customize pH buffer standard 

 

Up to 5 points calibration 

 

Auto-lock / automatic-endpoint detection function 

 

99 memory with real time clock recording 

 

RS232 online data logging to PC (sold separately) 

 
2.

 

  CONTENT: 

Carefully unpack the box.  It should contain the following items: 
a.

 

Main unit 

b.

 

Combination pH electrode 

c.

 

Temperature probe 

d.

 

Calibration buffer solution pH7.00 & pH4.01 of 90ml each. 

e.

 

Operations manual 

 

3.

 

SPECIFICATION: 

 

pH 

mV 

Temperature 

Range 

-2 to 16 pH  ±700.0mV   ±2,000mV 

0 to 120ºC 

Resolution 

0.01pH 

0.1mV  

1mV 

0.1ºC 

Accuracy 

±0.02pH 

±0.5mV 

±2mV 

±0.1 ºC 

±0.5ºC (Probe) 

Calibration 

Min. 2 point, Max. 5 point 

Buffer Recognition 

up to 60ºC or 95ºC depending on buffer type 

Automatic 
Temperature 
Compensation 

pH electrode temperature error compensation  

0 to 120ºC 

Memory 

99 data with real time clock 

Power source 

4 x 1.5Volt AAA size battery 

Operating 
temperature 

5 to 40ºC 

Storage temperature 

-20 to 60ºC 

Operating humidity 

Up to 95% RH 

Summary of Contents for WalkLAB HP 9010

Page 1: ...0 15 0 11 4 1 14 5 10 13 2 9 6 3 12 7 8 15 0 15 0 1 14 1 14 13 2 13 2 3 12 3 12 11 4 11 4 5 10 5 10 9 6 9 6 7 8 7 8 WalkLAB Professional pH Meter HP 9010 Operations Manual...

Page 2: ...2 online data logging to PC sold separately 2 CONTENT Carefully unpack the box It should contain the following items a Main unit b Combination pH electrode c Temperature probe d Calibration buffer sol...

Page 3: ...to snap down cover 4 PRODUCT LAYOUT pH Electrode Socket Temperature Probe Socket Battery Cover Lift up latch here to open Battery Cover On Off Button Press once to switch ON To switch OFF press and h...

Page 4: ...and Meter mS ppt ppm kPa ION DO Cond Salinity TDS S ATC Auto Lock AM PM HOLD pH Button keys legend correlate to button key Indicator will blink while the function mode is active Date or Time display L...

Page 5: ...tomatically sense a stabilized reading and locks the endpoint reading Factory preset is Yes Yes Yes Yes 1 Press the ENTER key to set display will show Yes Yes Yes Yes blinking 2 Press the SELECT key u...

Page 6: ...last calibration information about the electrode conditions A good pH electrode should have an Offset value no greater than 60mV and a Slope value of above 75 and below 115 Otherwise the electrode sh...

Page 7: ...ELECT key up or down to choose between C C C C and F F F F 3 Press the ENTER key to confirm 4 Press the SELECT key up for next setting Avto off AUTO OFF TIME SETTING This setting allow you set the tim...

Page 8: ...ENTER clock is up to date 5 Press the SELECT key up for next setting RST MASTER RESET This setting allows you to reset the unit to the original factory s default Original factory default Buffer Stand...

Page 9: ...current temperature Calibration with ISO Standard 7 5 Make sure the unit is properly installed and electrode connected 7 6 This meter can automatically recognize the ISO Standard buffer solution You w...

Page 10: ...bration cannot proceed until they are rectified 7 11 When the meter beeps Offset point is established 7 12 Now 1 68 4 01 10 01 and 12 45 will alternate on main display 7 13 Rinse the pH electrodes and...

Page 11: ...uffer solution Temperature C Slope 1 68 Slope 4 01 Offset 6 86 Slope 9 18 Slope 12 45 0 1 67 4 01 6 98 9 47 13 43 5 1 67 4 00 6 95 9 38 13 21 10 1 67 4 00 6 92 9 32 13 00 15 1 67 4 00 6 90 9 27 12 81...

Page 12: ...press EXIT for 2 seconds will abort all Calibration with Custom Standard 7 38 Make sure the unit is properly installed and electrode connected 7 39 Rinse the pH electrodes and temperature probe in dis...

Page 13: ...t reading is established You can now record the reading 8 5 Press the READ button for each new measurement Measurement with Auto Lock feature disabled see Page 4 8 6 When feature is disabled meter wil...

Page 14: ...Depend on the liquid that is measured some volatile liquid may take up to 20 minutes or even longer time to stabilize Disable the Auto lock Page4 and Auto Off Page6 features Then endpoint reading shou...

Page 15: ...asure liquid within specified range ErLc Prob Readings drift or unstable cannot auto lock during Read or Cal 1 Ground loop in solution 2 Electrode is clog or expired 3 Electromagnetic interference 4 O...

Page 16: ...nsinstruments com http www transinstruments com Singapore Pte Ltd email sales transinstruments com Website www transinstrumnets com ISO9001 Certified Firm Rev 4 Sep2015 PRINTED IN SINGAPORE Quality ch...

Reviews: