Trans instruments WalkLAB HP 9010 Operation Manual Download Page 12

 

11

 

7.26

 

If the meter does not recognize the buffer, it means the electrode 
could  be  defective  or  the  buffer  solution  is  wrong.    Calibration 
cannot proceed until they are rectified. 

7.27

 

When the meter beeps, Offset point is established. 

7.28

 

Now 1.68, 4.01, 9.18 and 12.45 will alternate on main display. 

7.29

 

Rinse the pH electrodes and temperature probe in distilled water, 
blot or shake dry then dip into next slope buffer solution. 

7.30

 

When the meter recognizes the buffer, it will appear blinking. 

7.31

 

If the value is different from the buffer solution then the electrode 
could be damage. 

7.32

 

Wait for meter to beep, Slope point is established. 

7.33

 

The remaining buffer values will alternate on the main display. 

7.34

 

Repeat step 7.29 to 7.33 sequentially up to 4 slope points. 

7.35 

At  least  2  calibration  points  (offset  &  slope)  must  be  established  in  each 
session of calibration.  Otherwise, calibrated point will not register.

 

7.36

 

When calibration is completed, press 

 ENTER 

 to store all calibrated 

points and return to normal reading mode. 

7.37

 

Anytime during calibration, press 

 EXIT 

 for 2 seconds will abort all. 

 

Calibration with Custom Standard 

7.38

 

Make sure the unit is properly installed and electrode connected. 

 

 

 

 

7.39

 

Rinse  the  pH  electrodes  and  temperature  probe  in  distilled  water 

and dip in the buffer solution. You will need at least two (2) buffers. 

Always begin with “Offset” calibration first

7.40

 

Press 

 CAL 

 and display will show 

7

7

7

7

.

00

00

00

00

 while 

 CAL 

 blinking, indicating 

it is in calibration mode. 

7.41

 

Press the 

 SELECT 

 button up / down to adjust the value. 

7.42

 

When the meter beeps, it means reading has stabilized and ready 
for calibration.  Press 

 ENTER 

 to confirm to calibrate to the point. 

7.43

 

A  preset  value  of 

1

1

1

1

.

40

40

40

40

3

3

3

3

.

80

80

80

80

and 

12

12

12

12

.

.60

.60

.60

.60

  will  now  alternate 

requesting for a slope value. 

7.44

 

Rinse the pH electrodes and temperature probe in distilled water, 
blot or shake dry then dip into next slope buffer solution. 

7.45

 

The meter will recognize the nearest buffer value and will appear 
blinking while it beeps in wait of a complete action. 

Always rinses the electrode with distilled water before and after each 

test.  This is to prevent solution carry over or cross contamination.  

Standard solutions must maintain highest purity; otherwise the 

meter’s accuracy could be compromised. 

Summary of Contents for WalkLAB HP 9010

Page 1: ...0 15 0 11 4 1 14 5 10 13 2 9 6 3 12 7 8 15 0 15 0 1 14 1 14 13 2 13 2 3 12 3 12 11 4 11 4 5 10 5 10 9 6 9 6 7 8 7 8 WalkLAB Professional pH Meter HP 9010 Operations Manual...

Page 2: ...2 online data logging to PC sold separately 2 CONTENT Carefully unpack the box It should contain the following items a Main unit b Combination pH electrode c Temperature probe d Calibration buffer sol...

Page 3: ...to snap down cover 4 PRODUCT LAYOUT pH Electrode Socket Temperature Probe Socket Battery Cover Lift up latch here to open Battery Cover On Off Button Press once to switch ON To switch OFF press and h...

Page 4: ...and Meter mS ppt ppm kPa ION DO Cond Salinity TDS S ATC Auto Lock AM PM HOLD pH Button keys legend correlate to button key Indicator will blink while the function mode is active Date or Time display L...

Page 5: ...tomatically sense a stabilized reading and locks the endpoint reading Factory preset is Yes Yes Yes Yes 1 Press the ENTER key to set display will show Yes Yes Yes Yes blinking 2 Press the SELECT key u...

Page 6: ...last calibration information about the electrode conditions A good pH electrode should have an Offset value no greater than 60mV and a Slope value of above 75 and below 115 Otherwise the electrode sh...

Page 7: ...ELECT key up or down to choose between C C C C and F F F F 3 Press the ENTER key to confirm 4 Press the SELECT key up for next setting Avto off AUTO OFF TIME SETTING This setting allow you set the tim...

Page 8: ...ENTER clock is up to date 5 Press the SELECT key up for next setting RST MASTER RESET This setting allows you to reset the unit to the original factory s default Original factory default Buffer Stand...

Page 9: ...current temperature Calibration with ISO Standard 7 5 Make sure the unit is properly installed and electrode connected 7 6 This meter can automatically recognize the ISO Standard buffer solution You w...

Page 10: ...bration cannot proceed until they are rectified 7 11 When the meter beeps Offset point is established 7 12 Now 1 68 4 01 10 01 and 12 45 will alternate on main display 7 13 Rinse the pH electrodes and...

Page 11: ...uffer solution Temperature C Slope 1 68 Slope 4 01 Offset 6 86 Slope 9 18 Slope 12 45 0 1 67 4 01 6 98 9 47 13 43 5 1 67 4 00 6 95 9 38 13 21 10 1 67 4 00 6 92 9 32 13 00 15 1 67 4 00 6 90 9 27 12 81...

Page 12: ...press EXIT for 2 seconds will abort all Calibration with Custom Standard 7 38 Make sure the unit is properly installed and electrode connected 7 39 Rinse the pH electrodes and temperature probe in dis...

Page 13: ...t reading is established You can now record the reading 8 5 Press the READ button for each new measurement Measurement with Auto Lock feature disabled see Page 4 8 6 When feature is disabled meter wil...

Page 14: ...Depend on the liquid that is measured some volatile liquid may take up to 20 minutes or even longer time to stabilize Disable the Auto lock Page4 and Auto Off Page6 features Then endpoint reading shou...

Page 15: ...asure liquid within specified range ErLc Prob Readings drift or unstable cannot auto lock during Read or Cal 1 Ground loop in solution 2 Electrode is clog or expired 3 Electromagnetic interference 4 O...

Page 16: ...nsinstruments com http www transinstruments com Singapore Pte Ltd email sales transinstruments com Website www transinstrumnets com ISO9001 Certified Firm Rev 4 Sep2015 PRINTED IN SINGAPORE Quality ch...

Reviews: