Trans instruments WalkLAB HP 9010 Operation Manual Download Page 13

 

12

 

7.46

 

Press the 

 

SELECT 

 

button up or down to adjust to the Slope value 

of the solution. 

7.47

 

Press 

  ENTER 

  to  confirm  set  value,  wait  a  while  for  an  endpoint 

reading to establish and press it again to calibrate to the point. 

7.48

 

The remaining slope values will alternate on the main display. 

7.49

 

Repeat step 7.43 to 7.47 in sequence.  You can calibrate up to 4 
slope points. 

7.50

 

At least 2 calibration point (offset & slope) must be established in each session 
of calibration.  Otherwise, calibrated point will not register.

 

7.51

 

When calibration is completed, press 

 EXIT 

 to resume measuring 

mode. 

MAKING PH MEASUREMENT: 

Measurement with Auto-Lock feature enabled:(see Page 4) 

8.1

 

Always  rinse  the  pH  electrode  and  temperature  probe  with 
distilled-water before and after each test. 

8.2

 

Dip  the  pH  electrode  and  temperature  probe  in  the  test  sample 
and press the 

 READ 

 button once. 

8.3

 

Display will show current pH reading with 

 AutoLock 

 symbol blinking. 

8.4

 

When the 

 AutoLock 

 sign stops with short beep, an endpoint reading 

is established.  You can now record the reading. 

8.5

 

Press the 

 READ 

 button for each new measurement. 

Measurement with Auto-Lock feature disabled: (see Page 4)

 

8.6

 

When feature is disabled, meter will display readings continuously. 

8.7

 

Press the 

 READ 

 button once to freeze displayed reading. 

8.8

 

The 

HOLD

  symbol  will  appear.    Press  again  will  release  display  and 

resume continuous measurement. 

 
 

 

 

 
 

pH measurement without ATC (Manual temperature compensation on sensor error) 

8.9

 

If  ATC  is  not  required,  disconnect  the  temperature  probe.   

ATC

 

symbol will not appear on screen and display will show the default 
temperature of 25ºC. 

8.10

 

You can also manually set the solution’s temperature as follow: 

8.11

 

Press the 

  READ 

 button start  measuring,  then 

press and hold down

 

 

MODE 

 

button until temperature reading appears blinking. 

Note on Measurements and Analysis: 

Each type of liquid has its own pH value varies at different temperature.  ATC 

measurement only compensates for the pH sensor’s error, it does not compensate 

for the pH variation due to temperature fluctuation in the liquid.  Therefore, each 

test should be made at the same temperature if possible or recorded with 

temperature reading for a comprehensive analysis. 

Summary of Contents for WalkLAB HP 9010

Page 1: ...0 15 0 11 4 1 14 5 10 13 2 9 6 3 12 7 8 15 0 15 0 1 14 1 14 13 2 13 2 3 12 3 12 11 4 11 4 5 10 5 10 9 6 9 6 7 8 7 8 WalkLAB Professional pH Meter HP 9010 Operations Manual...

Page 2: ...2 online data logging to PC sold separately 2 CONTENT Carefully unpack the box It should contain the following items a Main unit b Combination pH electrode c Temperature probe d Calibration buffer sol...

Page 3: ...to snap down cover 4 PRODUCT LAYOUT pH Electrode Socket Temperature Probe Socket Battery Cover Lift up latch here to open Battery Cover On Off Button Press once to switch ON To switch OFF press and h...

Page 4: ...and Meter mS ppt ppm kPa ION DO Cond Salinity TDS S ATC Auto Lock AM PM HOLD pH Button keys legend correlate to button key Indicator will blink while the function mode is active Date or Time display L...

Page 5: ...tomatically sense a stabilized reading and locks the endpoint reading Factory preset is Yes Yes Yes Yes 1 Press the ENTER key to set display will show Yes Yes Yes Yes blinking 2 Press the SELECT key u...

Page 6: ...last calibration information about the electrode conditions A good pH electrode should have an Offset value no greater than 60mV and a Slope value of above 75 and below 115 Otherwise the electrode sh...

Page 7: ...ELECT key up or down to choose between C C C C and F F F F 3 Press the ENTER key to confirm 4 Press the SELECT key up for next setting Avto off AUTO OFF TIME SETTING This setting allow you set the tim...

Page 8: ...ENTER clock is up to date 5 Press the SELECT key up for next setting RST MASTER RESET This setting allows you to reset the unit to the original factory s default Original factory default Buffer Stand...

Page 9: ...current temperature Calibration with ISO Standard 7 5 Make sure the unit is properly installed and electrode connected 7 6 This meter can automatically recognize the ISO Standard buffer solution You w...

Page 10: ...bration cannot proceed until they are rectified 7 11 When the meter beeps Offset point is established 7 12 Now 1 68 4 01 10 01 and 12 45 will alternate on main display 7 13 Rinse the pH electrodes and...

Page 11: ...uffer solution Temperature C Slope 1 68 Slope 4 01 Offset 6 86 Slope 9 18 Slope 12 45 0 1 67 4 01 6 98 9 47 13 43 5 1 67 4 00 6 95 9 38 13 21 10 1 67 4 00 6 92 9 32 13 00 15 1 67 4 00 6 90 9 27 12 81...

Page 12: ...press EXIT for 2 seconds will abort all Calibration with Custom Standard 7 38 Make sure the unit is properly installed and electrode connected 7 39 Rinse the pH electrodes and temperature probe in dis...

Page 13: ...t reading is established You can now record the reading 8 5 Press the READ button for each new measurement Measurement with Auto Lock feature disabled see Page 4 8 6 When feature is disabled meter wil...

Page 14: ...Depend on the liquid that is measured some volatile liquid may take up to 20 minutes or even longer time to stabilize Disable the Auto lock Page4 and Auto Off Page6 features Then endpoint reading shou...

Page 15: ...asure liquid within specified range ErLc Prob Readings drift or unstable cannot auto lock during Read or Cal 1 Ground loop in solution 2 Electrode is clog or expired 3 Electromagnetic interference 4 O...

Page 16: ...nsinstruments com http www transinstruments com Singapore Pte Ltd email sales transinstruments com Website www transinstrumnets com ISO9001 Certified Firm Rev 4 Sep2015 PRINTED IN SINGAPORE Quality ch...

Reviews: