Trans instruments WalkLAB HP 9010 Operation Manual Download Page 6

 

5

 

Bvf

CALIBATION BUFFER STANDARDS 

This setting allows you to select ISO, NIST or customized 
buffer standard.  Factory preset is ISO standard. 

ISO

  (

I

I

I

ISO

SO

SO

SO

)

buffer  standards  are  pH1.68,  7.00,  4.01,  10.01, 

12.45  and 

NIST

  (

I

7

I

7

I

7

I

7

   I

   I

   I

   ISt

St

St

St

)

  buffer  standards  are  pH1.68, 

4.01,  6.86,  9.18  and  12.45,  which  are  preset  in  this 
meter,  and  will  be  automatically  recognized  during 
calibration. 
Select 

customize

  (

CvSt

CvSt

CvSt

CvSt

if  you  have  other  standards 

other than standard buffer solutions. 

1. 

Press the 

 ENTER 

 

key to set; display will show 

I

I

I

ISO

SO

SO

SO

 

blinking.  

2.

 

Press the 

 SELECT 

 key 

up

 or 

down 

to select from 

I

I

I

ISO

SO

SO

SO

,  

I

7

I

7

I

7

I

7

   I

   I

   I

   Ist

st

st

st

 or 

CvSt

CvSt

CvSt

CvSt

 and press the 

 ENTER 

 key to confirm. 

3. 

Press the 

SELECT 

 key 

up 

for next setting.

 

 

stAt

PrB

PrB

PrB

PrB

PROBE ELECTRODE STATUS

 

This  setting  allows  you  to  review  the  last  calibration 
information about the electrode conditions. 

A good pH electrode should have an Offset value no 
greater than ±60mV and a Slope value of above 75% 
and below 115%.  Otherwise, the electrode should be 
replaced.

 

There are 5 calibration points for review as follow: 

 

Displayed Value  Calibration Point within range 

OFF

OFF

OFF

OFF 

 

 

 

Offset in mV 

pH7.00 

SL1

SL1

SL1

SL1 

 

 

 

% of slope 

pH 7.00 to 4.01 

SL2

SL2

SL2

SL2 

 

 

 

% of slope 

pH 7.00 to 10.01 

SL3

SL3

SL3

SL3 

 

 

 

% of slope 

pH 4.01 to -2.00 

SL4

SL4

SL4

SL4 

 

 

 

% of slope 

pH 10.01 to 16.00 

1.

 

Press the 

 ENTER 

 key to set; display will show the mV 

value with

 

Off

Off

Off

Off

 blinking.  

2.

 

Press  the 

  SELECT 

  key 

up

  to  review  Slope  point  1, 

display will show the % of slope with 

sl1

sl1

sl1

sl1

 below. 

3.

 

Press  the 

  SELECT 

  key 

up

  sequentially  to  review  all 

the  rest  of  slope  value  2,  3,  4  and  then  back  to 
Offset in a cyclical mode. 

4. 

When  finish,  press  the 

  EXIT 

  key  once  followed  by 

the 

SELECT 

 key 

up 

for next setting.

 

 

 

IsO

 

 

    

bvf

 

 

       

SETUP 

ENTER 

 

SELECT 

EXIT 

 
                                %

 

     98.5

 

 

   

sl2

 

 

       

 

stAt

 

 

    

Prb

 

 

       

SETUP 

ENTER 

 

SELECT 

EXIT 

 

      

0.5

  mV

 

 

 

    

Off

 

 

       

SETUP 

ENTER 

 

SELECT 

EXIT 

SETUP 

ENTER 

 

SELECT 

EXIT 

Summary of Contents for WalkLAB HP 9010

Page 1: ...0 15 0 11 4 1 14 5 10 13 2 9 6 3 12 7 8 15 0 15 0 1 14 1 14 13 2 13 2 3 12 3 12 11 4 11 4 5 10 5 10 9 6 9 6 7 8 7 8 WalkLAB Professional pH Meter HP 9010 Operations Manual...

Page 2: ...2 online data logging to PC sold separately 2 CONTENT Carefully unpack the box It should contain the following items a Main unit b Combination pH electrode c Temperature probe d Calibration buffer sol...

Page 3: ...to snap down cover 4 PRODUCT LAYOUT pH Electrode Socket Temperature Probe Socket Battery Cover Lift up latch here to open Battery Cover On Off Button Press once to switch ON To switch OFF press and h...

Page 4: ...and Meter mS ppt ppm kPa ION DO Cond Salinity TDS S ATC Auto Lock AM PM HOLD pH Button keys legend correlate to button key Indicator will blink while the function mode is active Date or Time display L...

Page 5: ...tomatically sense a stabilized reading and locks the endpoint reading Factory preset is Yes Yes Yes Yes 1 Press the ENTER key to set display will show Yes Yes Yes Yes blinking 2 Press the SELECT key u...

Page 6: ...last calibration information about the electrode conditions A good pH electrode should have an Offset value no greater than 60mV and a Slope value of above 75 and below 115 Otherwise the electrode sh...

Page 7: ...ELECT key up or down to choose between C C C C and F F F F 3 Press the ENTER key to confirm 4 Press the SELECT key up for next setting Avto off AUTO OFF TIME SETTING This setting allow you set the tim...

Page 8: ...ENTER clock is up to date 5 Press the SELECT key up for next setting RST MASTER RESET This setting allows you to reset the unit to the original factory s default Original factory default Buffer Stand...

Page 9: ...current temperature Calibration with ISO Standard 7 5 Make sure the unit is properly installed and electrode connected 7 6 This meter can automatically recognize the ISO Standard buffer solution You w...

Page 10: ...bration cannot proceed until they are rectified 7 11 When the meter beeps Offset point is established 7 12 Now 1 68 4 01 10 01 and 12 45 will alternate on main display 7 13 Rinse the pH electrodes and...

Page 11: ...uffer solution Temperature C Slope 1 68 Slope 4 01 Offset 6 86 Slope 9 18 Slope 12 45 0 1 67 4 01 6 98 9 47 13 43 5 1 67 4 00 6 95 9 38 13 21 10 1 67 4 00 6 92 9 32 13 00 15 1 67 4 00 6 90 9 27 12 81...

Page 12: ...press EXIT for 2 seconds will abort all Calibration with Custom Standard 7 38 Make sure the unit is properly installed and electrode connected 7 39 Rinse the pH electrodes and temperature probe in dis...

Page 13: ...t reading is established You can now record the reading 8 5 Press the READ button for each new measurement Measurement with Auto Lock feature disabled see Page 4 8 6 When feature is disabled meter wil...

Page 14: ...Depend on the liquid that is measured some volatile liquid may take up to 20 minutes or even longer time to stabilize Disable the Auto lock Page4 and Auto Off Page6 features Then endpoint reading shou...

Page 15: ...asure liquid within specified range ErLc Prob Readings drift or unstable cannot auto lock during Read or Cal 1 Ground loop in solution 2 Electrode is clog or expired 3 Electromagnetic interference 4 O...

Page 16: ...nsinstruments com http www transinstruments com Singapore Pte Ltd email sales transinstruments com Website www transinstrumnets com ISO9001 Certified Firm Rev 4 Sep2015 PRINTED IN SINGAPORE Quality ch...

Reviews: