3.3.4 44°C Program (Fig1.3)
This mode is the pre-programmed default 44°C incubation cycle. In this mode the incubator will
run a 24 hour incubation cycle at 44°C. The timer can be altered to any setting between 0 and 24
hours. Adjusting the incubation time is described in section 3.3.7.
Pressing and holding the middle switch for 5 seconds will start the incubation cycle and the screen
will change to the following:
The asterisk preceding the temperature will switch on and off indicating the status of the heater.
To stop or pause the incubation cycle press and hold the middle switch again for 5 seconds. The
display will return to the one shown in Fig 1.2.
20.9°C 24:00
->44° MEM
10.5v
*20.9°C 24:00
->44° RUN! 10.5v