
Page 59
22.3 pH Temperature Compensation
The slope of a pH sensor (section 22.2) is affected by temperature. This
effect is compensated for either by using an Automatic Temperature
Compensation (ATC) probe or by entering the sample temperature
manually. Figure 22-3 shows the slope of a pH sensor at various
temperatures.
-600
-400
-200
0
200
400
600
0
7
14
Millivolts
Electrode
Potential (mV) at
0 oC (54mV/pH)
Electrode
Potential (mV) at
50 oC (64mV/pH)
Electrode
Potential (mV) at
100 oC
(74mV/pH)
pH Electrode Response, as a Function of Temperature
pH
Figure 22-3
22.4 Checking the reference junction of a pH sensor
If pH readings are inaccurate or unstable, the reference junction of the
sensor may be blocked. The following test can be performed to
determine if the reference junction of a pH sensor is making adequate
contact with the sample solution.
1. Calibrate the
WP-91
, as per section 5.
2. Dilute 1 part of pH6.88 buffer with 9 parts of distilled water.
3. Measure the pH of the diluted buffer. The result should be 7.06 +/-
0.02 pH.
4. If the value obtained is outside of these limits, then clean the
reference junction, as per the instructions supplied with the pH
sensor.
5. Re-calibrate the
WP-91
and repeat the test.
6. If the value obtained is still outside 7.06 +/-0.02 pH, then the sensor
should be replaced.