pH and mV Troubleshooting, continued…
Inaccurate
readings, even
when calibration
is successful.
Reference junction blocked.
Clean reference junction, as
per instructions supplied
with the sensor.
Displays 7.00 for
all solutions.
Electrical short in connector.
1. Check connector.
Replace if necessary.
2. Replace sensor.
Displays 4-5 pH
for all solutions.
Glass bulb or internal stem
cracked.
Replace sensor.
20.4 Temperature Troubleshooting
Symptom
Possible Causes
Remedy
Displays “
OVR
o
C
”
when Dissolved
Oxygen or
Temperature
sensor is plugged
in.
1. Faulty sensor.
2. Faulty instrument.
Return sensor to factory for
repair or replacement.
Return to factory for repair.
Temperature
inaccurate and
cannot be
calibrated.
1. Faulty connector.
2. Faulty Dissolved Oxygen
or Temperature sensor.
3. Faulty instrument.
Check the connector and
replace if necessary.
Return sensor to factory for
repair or replacement.
Return to factory for repair.