
Page 50
20.3 pH and mV Troubleshooting
Symptom
Possible Causes
Remedy
Unit fails to
calibrate, even
with new probe.
Calibration settings outside
of allowable limits due to
previous failed calibration.
Initialise the unit. See
section 18.
1 Point calibration
fails (Asymmetry
is greater than +/-
1.00 pH).
1. Reference junction
blocked.
2. Reference electrolyte
contaminated.
Clean reference junction, as
per instructions supplied
with the sensor.
Flush with distilled water
and replace electrolyte.
2 Point calibration
fails (Slope is less
than 85.0%).
1. Incorrect primary buffer.
2. Glass bulb not clean.
3. Sensor is aged.
4. Connector is damp.
5. Buffers are inaccurate.
Ensure that you are using
the primary buffer for which
the
WP-91
has been set (see
section 17).
Clean glass bulb as per
instructions supplied with
the sensor.
Attempt rejuvenation, as per
instructions supplied with
the sensor. If not successful,
replace sensor.
Dry in a warm place.
Replace buffers.
Unstable
readings.
1. Electrolyte chamber
needs to be refilled.
2. Reference junction
blocked.
3. Glass bulb not clean.
4. Bubble in glass bulb.
5. Faulty connection to
meter.
6. Reference junction not
immersed.
7. KCl crystals around
reference junction, inside
the electrolyte chamber.
Refill with saturated KCl
filling solution.
Clean reference junction, as
per instructions supplied
with the sensor.
Clean glass bulb as per
instructions supplied with
the sensor.
Flick the sensor to remove
bubble.
Check connectors. Replace
if necessary.
Ensure that the bulb AND
the reference junction are
fully immersed.
Rinse electrolyte chamber
with warm distilled water
until dissolved. Replace
electrolyte.
Continued next page…