Page 47
18. Initialising the WP-91
If the calibration settings of the
WP-91
exceed the allowable limits, the
unit may need to be initialised to factory default values.
To initialise the
WP-91
:
1. Switch the
WP-91
OFF.
2. Press and hold the
key while switching the
WP-91
back ON.
3. The following messages are displayed…
Initialized
MUST ReCalibrate
then:
WP91s V1.2 S1234
Oxy. pH/mV Temp.
(The “s” after
WP-91
is shown when the RS232 serial port option is
fitted)
4. The meter then displays Dissolved Oxygen, pH and Temperature.
Note that the decimal points have been replaced with a
∗
, to indicate
that the unit requires recalibration.
5. When the
WP-91
is initialised…
(a) The manual salinity correction value is re-set 36.0ppK. See
section 8 if you wish to change this value.
(b) The Altitude or Pressure correction system is switched off, and
the instrument assumes sea level (0m) and 1013HPa.
(c) The automatically recognised buffers are re-set to pH4.00,
pH6.88 and pH9.23. To change these, see section 17.
19. Instrument firmware version number.
If you need to phone or fax TPS for any further technical assistance, the
version number of your
WP-91
firmware may of benefit to us. The
version number is displayed by the
WP-91
at turn-on.