7.2 Temperature Calibration Notes
1.
Temperature calibration information is retained in memory when the
smart
CHEM
-Titro
is
switched off, even when the power supply is removed. This information can be recalled later
using the GLP function (see section 8).
2.
Temperature does not need to be re-calibrated unless the Temperature sensor is replaced or the
meter is initialised.
7.3 Calibration Messages
1.
If a temperature calibration has been successfully performed, the
smart
CHEM
-Titro
will
display the following message and the offset of the sensor.
Calibration OK
Offset=0.1
O
C
2.
If a temperature calibration has failed, the
smart
CHEM
-Titro
will display the following
message and the failed offset value of the sensor.
Calibration Failed
Offset=11.0
O
C
3.
The
smart
CHEM
-Titro
has an allowable Offset range of -10.0 to +10.0
O
C
. If calibration fails
due to the Offset being outside these limits, then please consult the Troubleshooting guide
(section 17.3) for possible remedies.