Page 20
7. Recorder Output Option
This section is applicable when the optional analogue recorder output is fitted.
The recorder output depends on the currently selected display mode, as detailed
below…
Display Mode
Parameter sent to Recorder Port
Conductivity & Temperature
Conductivity
TDS and Temperature
TDS
Temperature only
Temperature
RS232 Send Rate
No Output
RS232 Baud Rate
No Output
The output voltages are as follows:
Cond................0 to 2000 Counts for 0 to 2000 mV
ie. 2.76 mS/cm = 276 mV
TDS.................0 to 1000 Counts for 0 to 1000 mV
ie. 36.0 ppK = 360 mV
Temperature ....-10.0 to 120.0
O
C for 0 to 2000 mV
i.e. 0.0
O
C = 152 mV Output
Output impedance approx 1000 Ohms.