3.2 Calibration Notes
1.
A Zero calibration should be performed at least monthly. In low conductivity
applications (where a zero error is particularly significant), a zero calibration
may have to be done weekly.
2.
A Standard calibration should be performed at least weekly. Of course, more
frequent calibration will result in greater confidence in results.
3.
Conductivity and TDS calibration data is stored separately in memory. Ensure
that the
901-C
has been correctly calibrated for the mode in which it will be
used. The
901-C
does not require re-calibration when alternating between
Conductivity and TDS modes, providing the instrument has been correctly
calibrated for both.
4.
All calibration information is retained in memory when the
901-C
is switched
off, even when the power supply is removed.