
Page 14
4.3 Calibration Messages
1.
If a Zero calibration has been successfully performed, the
901-C
will display
the Zero of the sensor and then return to TDS mode. For example…
Zero OK, 0.1%
Zero OK, 0.1%
2.
If a Span calibration has been successfully performed, the
901-C
will
display the k factor of the sensor and then return to TDS mode. For
example…
Cal OK, k=1.00
Cal OK, k=1.00
Note that "
∗∗
" in the TDS reading has now been replaced by a decimal
point, due to the successful calibration.
3.
If a Span calibration has failed, the
901-C
will display the following
messages and then the failed k factor of the sensor before returning to TDS
mode. For example…
Calibrate Failed
Calibrate Failed
then:
STD=36.00ppK ?
STD=36.00ppK ?
then:
k=0.70, Fails
k=0.70, Fails
Note that a "
∗∗
" replaces the decimal point in the TDS reading to indicate
that TDS is not correctly calibrated.
Check that the sensor is immersed at least to the vent hole in the body, and
that the standard is correct before attempting calibration again.