48
18. Troubleshooting
18.1 General Errors
Error Message
Possible Causes
Remedy
Factory Calibration
Data Failure
The EEPROM chip which
contains the factory calibration
information has failed.
The unit must be returned to TPS for
service.
EEPROM Write
Failure
Return to Factory
for Service
User calibration settings have
been lost or corrupted.
Switch the meter OFF and switch back ON.
If the problem persists, return the unit to
TPS for service.
Flashing
symbol.
Battery is below 5.60 volts.
Recharge the battery. A full charge will
take approximately 18 hours. Note that the
unit will switch itself off when the battery
falls below 5.10 volts.
Meter displays the word
OFF
, and switches off.
Battery is below 5.10 volts.
Recharge the battery.
If this fails, check the charger.
If charger is OK, replace the battery.
Meter will not turn on.
Battery is exhausted.
Recharge the battery for approximately 18
hours.
If this fails, check the charger.
If charger is OK, replace the battery.
Battery does not charge up
when charger is connected.
1.
Faulty battery charger.
2.
Faulty battery.
1.
Connect the charger and switch the
power on.
2.
Display the battery volts in the battery
saver menu (see section 16).
3.
If the battery volts are increasing then
the charger is OK. If the battery volts
do not increase, then the charger is
faulty.
4.
Replace the charger or the battery, as
required.