3.2.2 Span Calibration in Air (all Oxygen modes)
1.
Hang the Dissolved Oxygen sensor in air. The tip of the Dissolved Oxygen sensor should be
pointing downwards.
Allow the reading to stabilise. After a zero calibration, this may take up to 5 minutes.
2.
Select Oxygen Calibration. (
→
→
→
→
F1:Calibrate
→
→
→
→
F1:Oxygen
)
When the reading is above approximately 25% Saturation, 2 ppM or 5% Gaseous, the
90-FLMV
will display the AIR calibration screen…
101.0%Sat 25.0
o
c
Oxygen AIR Calibration, Press Enter
Press
to calibrate.
A “
∗
” in the display will be replaced by a decimal point after a successful air calibration.
3.
The
90-FLMV
is now calibrated and is ready for Dissolved Oxygen measurements. Rinse the
Dissolved Oxygen sensor in distilled water and blot dry before placing it into unknown
samples.
3.2.3 Span Calibration in Solution (Salinity-corrected ppM Mode only)
This span calibration provides an alternative to calibrating the Dissolved Oxygen sensor in air. It is
only available when the
90-FLMV
is in Salinity-corrected ppM mode. Please note that the normal
AIR calibration (section 3.2.2) is still available for Salinity-corrected ppM mode.
1.
Measure the Dissolved Oxygen content of the solution to be used for calibration. This is
generally done with a Winkler titration. The
90-FLMV
span calibration should be performed
immediately the Dissolved Oxygen content of the solution is known, as the value may not be
stable.
2.
Place the Dissolved Oxygen and Conductivity sensors into the calibration solution. Ensure that
the Conductivity sensor is calibrated and is correctly immersed (see section 4.4).
The solution must be stirred at a moderate rate.
Allow the reading to stabilise. After a zero calibration, this may take up to 5 minutes.
3.
Select Oxygen Calibration. (
→
→
→
→
F1:Calibrate
→
→
→
→
F1:Oxygen
)
When the reading is above approximately 2 ppM, the
90-FLMV
will display the AIR/SPAN
calibration screen. Note the cursor underlining the “
A
” in “
Air
”.
9.10ppM 25.0
o
c
Oxygen AIR/SPAN Calibration, Press Enter
Use the numeric keypad to enter the Dissolved Oxygen value of the solution. The words
“AIR/SPAN” are deleted and the value being entered is displayed. The Conductivity or TDS or
Salinity reading also appears as soon as the first numeric key is pressed.
Press the
to correct any errors.
Ensure that the Dissolved Oxygen, Conductivity or TDS or Salinity, and Temperature readings
are fully stable.
Press
to calibrate.
A “
∗
” in the display will be replaced by a decimal point after a successful air calibration.
4.
The
90-FLMV
is now calibrated and is ready for Dissolved Oxygen measurements. Rinse the
Dissolved Oxygen sensor in distilled water and blot dry before placing it into unknown
samples.