3.5 Dissolved Oxygen Stirrer
The
90-FLMV
is equipped with a 4.5V DC output to power a stirrer for the Dissolved Oxygen
sensor. This power output is suitable for the TPS submersible DO
2
stirrer (part number 123306).
3.5.1 Enabling and Disabling the Dissolved Oxygen stirrer output
1.
Select the System menu (
→
→
→
→
F5:System
).
2.
Select
F4:Stirrer Enable
or
F4:Stirrer Disable
from the menu as required.
3.5.2 Connecting the Dissolved Oxygen stirrer
To connect the Dissolved Oxygen stirrer…
1.
Plug the Dissolved Oxygen stirrer into the
Oxygen
socket on the meter.
2.
Plug the Dissolved Oxygen sensor into the in-line socket that is provided on the stirrer cable.
Please refer to the diagram below.
3.5.3 Starting and Stopping the Dissolved Oxygen stirrer
1.
Ensure that the Dissolved Oxygen stirrer output has been enabled, as per section 3.5.1.
2.
Press
once in normal display mode to start the stirrer. The stirrer will stay on for 40 seconds
and then stop. A countdown is provided on the screen.
3.
Press
a second time any time during the 40 second period to set the Dissolved Oxygen
stirrer to operate continuously.
4.
Press
a third time to stop the Dissolved Oxygen stirrer.
The Dissolved Oxygen stirrer starts and stops automatically when the 90-FLMV is in Rate per Day
automatic data logging mode. See section 11.3.1.