![TPS 90-FLMV Manual Download Page 17](http://html.mh-extra.com/html/tps/90-flmv/90-flmv_manual_1146433017.webp)
17
9.
When the reading has stabilised, press
to calibrate. The “
∗
” will now be replaced by a
decimal point if calibration was successful.
10.
The
90-FLMV
is now calibrated for Conductivity. The TDS and Salinity values are derived
from the Conductivity reading and do not require a separate calibration.
Ensure that the sensor is immersed at least as deeply as per the diagram in step 7 for all sample
measurements.
4.5 Conductivity Calibration Notes
1.
A Zero calibration should be performed at least monthly. In low conductivity applications
(where a zero error is particularly significant), a zero calibration may have to be done weekly.
2.
A Standard calibration should be performed at least weekly. Of course, more frequent
calibration will result in greater confidence in results.
3.
All calibration information is retained in memory when the
90-FLMV
is switched off. This
information can be recalled or printed later using the GLP function (see section 10).
4.
The
90-FLMV
displays the value of the standard to which it will attempt to calibrate. Ensure
that the standard value displayed corresponds to the standard that you are using. Alter the
Standards set-up if necessary (see section 4.2).
5.
Calibration settings for k=0.1, k=1 and k=10 sensors are NOT stored separately.
The
90-FLMV
requires re-calibration when a new k factor sensor is connected.
4.6 Conductivity Calibration Messages
1.
If a Zero Calibration has been successfully performed, the
90-FLMV
will display the following
message…
0.00uS 25.0
o
c
Calibration OK, Zero=0.01uS
2.
If a Standard Calibration has been successfully performed, the
90-FLMV
will display the
following message and the calculated k factor of the sensor. For example…
2.76mS 25.0
o
c
Calibration OK, k=0.99
3.
If a Standard Calibration has failed, the
90-FLMV
will display the following message and the
calculated k factor of the sensor. For example…
Calibration Failure. Check STD=2760uS/cm
k=3.64, Exceeds Limit
Notes
1.
The allowable k factor range is +/-25% of nominal. This range is ample to allow for correctly
functioning Conductivity sensors. If calibration fails due to the k factor being outside these
limits, then please consult the Troubleshooting guide (section 18.3) for possible remedies.