Page 25
10.3.2 ED1M Dissolved Oxygen Sensor Configuration
1. Set the
YSI or ED1M (J5)
jumper in the terminal area to be open. Fit the
jumper to one of the pins, as shown below, in case this is needed later.
J5
YSI
or
ED1
2. (a) Set the
YSI or ED1M (J3)
jumper in the terminal area across the first
jumper (from the left).
YSI
or
ED1M
J3
(b) When the
mini
CHEM
-DO
2
is set to ppM readout and the samples have a
high salinity, there may not be enough adjustment available with the
SPAN
control to calibrate the unit. If this is the case, set the
YSI or
ED1M (J3)
jumper to the following setting…
YSI
or
ED1M
J3