TPC PC2800 Operation Manual Download Page 6

 

Radio frequency 

emission 

GB4824 

 

frequency emission is very low, and the interference to 

nearby electronic equipment may be very small. 

Harmonic emission 

GB 17625.1 

Not applicable 

Dental  electric  suction  machine  is  suitable  for  non-

domestic  use  and  all  facilities  that  are  not  directly 

connected to the public low-voltage power supply network 

of domestic residences. 

Voltage fluctuation / 

flicker 

GB 17625.2 

 

Not applicable 

 

Guidance and manufacturer's declaration-electromagnetic immunity 

Dental electric suction machine is used in the electromagnetic environment specified below. The user or user 

who purchased this product should ensure that it is used in this electromagnetic environment. 

Immunity test 

IEC 60601 Electrical 

Level 

Coincidence Electrical 

level 

Electromagnetic 

environment-guide 

Electrostatic 

discharge (ESD) 

GB/T17626.2 

±

6 kV Contact 

discharge 

±

8 kV Air discharge 

±

6 kV Contact 

discharge 

±

8 kV Air discharge 

The ground should be wood, 

concrete or ceramic tiles. If the 

ground is covered with synthetic 

materials, the relative humidity 

should be at least 30%. 

Electrical  fast 

transient burst 

GB/T17626.4 

±

2kV On the power 

cord 

±

1kV For input / 

output lines 

±

2kV On the power 

cord 

Not Applicable 

The grid power supply should 

have the quality used in a typical 

commercial or hospital 

environment. 

Surge GB/T17626.5 

±

1 kV Line-to-Line 

±

2 kV Line-to-ground 

±

1 kV Line-to-Line 

±

2 kV Line-to-ground 

The grid power supply should 

have the quality used in a typical 

commercial or hospital 

environment. 

Voltage dips, short 

interruptions and 

voltage changes on 

the power input line 

GB/T17626.11 

<5% UT for 0.5 cycle 

(> 95% drop in UT) 

40% UT for 5 cycles 

(On UT, 60% drop) 

70% UT for 25 cycles 

(At UT,> 30% drop) 

<5% UT for 5 S 

(> 95% drop in UT) 

<5% UT for 0.5 cycle 

(> 95% drop in UT) 

40% UT for 5 cycles 

(On UT, 60% drop) 

70% UT for 25 cycles 

(At UT,> 30% drop) 

<5% UT for 5 S 

(> 95% drop in UT) 

The network power supply 

should have the quality used in a 

typical commercial or hospital 

environment. If the user of the 

dental electric suction machine 

needs to continue to operate 

during the power interruption, it 

is recommended that the dental 

electric suction machine be 

powered by an uninterruptible 

power supply or battery. 

Power 

frequency 

magnetic field 

GB/T17626.8 

3A/m 

3A/m 

The power frequency magnetic 

field should have the 

characteristics of power 

frequency magnetic field level in 

a typical place in a typical 

commercial or hospital 

Summary of Contents for PC2800

Page 1: ...ProClean Extra Oral Suction System Model PC2800 TPC 851 Lawson Street City of Industry Ca 91748 P 800 560 8222 www tpcdental com...

Page 2: ...1 Contents 1 Overview 2 2 Technical Specifications 4 3 Product Structure and Installation 6 4 Maintenance 8 5 Trouble shooting 8 6 Electric principle diagram 8 7 Disposing notes 9 8 Packing list 9...

Page 3: ...shock Before powering on this product please confirm that the power supply you provide can meet the input power information marked near the power input port of this product Do not share the socket wit...

Page 4: ...id the use of other highly sensitive instruments and equipment at the same time so as to avoid the equipment from being interfered by electromagnetic sources 1 4 4 Cleaning Please keep the environment...

Page 5: ...air 2 4 Electromagnetic Compatibility Medical electrical equipment requires special tips on electromagnetic compatibility as well as installation and use according to the electromagnetic compatibilit...

Page 6: ...t burst GB T17626 4 2kV On the power cord 1kV For input output lines 2kV On the power cord Not Applicable The grid power supply should have the quality used in a typical commercial or hospital environ...

Page 7: ...joint at the top of the box 3 Install the suction cover onto designated spot as above picture shows 4 Use one phase AC 220V 50hz or other local power 220V 60hz 240V 50hz 110V 60hz would beok 3 3 Oper...

Page 8: ...can effectively remove the droplets and aerosolparticles absorb water vapor Secondary Filter HEPA 13 filter it can effectively filter out 99 95 solid particles and colloidal substances Size 240 175 2...

Page 9: ...lamp once a year 4 4 ProClean Extra Oral Suction System When not in use store in a dry and ventilated place 4 5 Replacement Filters Items PC2800 SYNFLTR Replacement washable synthetic fiber filter PC...

Page 10: ...on cover pc 2 3 Instruction manual pc 1 7 Disposing notes Following requirements shall be followed as stored Disconnect the appliance from power The components of the product are non toxic dispose of...