TPC PC2800 Operation Manual Download Page 3

1.

 

Overview 

Thank you very much for choosing to the ProClean Extra Oral Suction System. Please read this instruction 

manual and the accompanying materials in detail.   

The ProClean is covered by a 

2 year warranty

 (not including replacement filters and uv lamp) 

1.1

 

Contents 

This instruction manual includes an overview, instructions for use of the suction unit, operation routines, 

maintenance, and precautions. You can view related content as needed and get corresponding help. 

1.2

 

Structure and Composition 

The external-oral suction unit is mainly composed of a generally vacuum pressure suction module, a 

primary filter, high-efficiency filters, ultraviolet sterilization, plasma sterilization, and a free arm. 

1.3

 

Ideal Applications 

It is used for the suction and purification of bacterial aerosols between medical staff and patients during 

dental treatment. 

1.4

 

Safety Information 

1.4.1

 

Electrical Safety 

 

 

The  power  supply  for  the  environment  in  which  this  product  is  used  must  have  a  good,  standard 

grounding  device,  and  the  grounding  wire  must  be  firm.  It  must  use  the  national  standard  single-phase 

three-plug power cord, and cooperate with a well-grounded single-phase three-hole socket to ensure good 

grounding of this product. 

   

Attention

To avoid the risk of electric shock, the device can only be connected to a power cable 

with a protective ground. An incorrect grounding plug can cause a risk of electric shock. 

 

Before powering on this product, please confirm that the power supply you provide can meet the 

input power information marked near the power input port of this product. 

 

Do not share the socket with other electrical appliances to prevent the voltage from being unstable 

sometimes, which may damage the product 

 

When performing maintenance / repair or cleaning this product, be sure to unplug the power cord to 

ensure that the product is completely powered off before proceeding. 

 

Regularly check the power cord and power plug for damage, and make sure that the power cord is 

not crushed by other objects. 

1.4.2

 

Regular Inspection 

Regular safety inspections should be performed by trained and skilled personnel, the following items 

should be checked at least once a year: 

Check the functional status of the equipment and machines. 

Check whether the relevant safety signs are clear. 

Check whether the performance of the device meets the instructions in the manual. 

Test whether the grounding resistance is less than the resistance value 0.1

 required in 

GB9706.1

1.4.3

 

Precautions 

1)

 

You should read and understand all contents in the manual carefully before operating. 

2)

 

Follow all warnings and instructions on the instrument during operation. 

3)

 

During maintenance and repair, the water and electricity sources must be emptied before the water and 

electricity sources are turned off. 

4)

 

Only trained and skilled medical personnel can operate the equipment. Improper use will cause serious 

personal injury. 

5)

 

In following cases, please do not use the device, nor can you open the device or replace the accessories 

Summary of Contents for PC2800

Page 1: ...ProClean Extra Oral Suction System Model PC2800 TPC 851 Lawson Street City of Industry Ca 91748 P 800 560 8222 www tpcdental com...

Page 2: ...1 Contents 1 Overview 2 2 Technical Specifications 4 3 Product Structure and Installation 6 4 Maintenance 8 5 Trouble shooting 8 6 Electric principle diagram 8 7 Disposing notes 9 8 Packing list 9...

Page 3: ...shock Before powering on this product please confirm that the power supply you provide can meet the input power information marked near the power input port of this product Do not share the socket wit...

Page 4: ...id the use of other highly sensitive instruments and equipment at the same time so as to avoid the equipment from being interfered by electromagnetic sources 1 4 4 Cleaning Please keep the environment...

Page 5: ...air 2 4 Electromagnetic Compatibility Medical electrical equipment requires special tips on electromagnetic compatibility as well as installation and use according to the electromagnetic compatibilit...

Page 6: ...t burst GB T17626 4 2kV On the power cord 1kV For input output lines 2kV On the power cord Not Applicable The grid power supply should have the quality used in a typical commercial or hospital environ...

Page 7: ...joint at the top of the box 3 Install the suction cover onto designated spot as above picture shows 4 Use one phase AC 220V 50hz or other local power 220V 60hz 240V 50hz 110V 60hz would beok 3 3 Oper...

Page 8: ...can effectively remove the droplets and aerosolparticles absorb water vapor Secondary Filter HEPA 13 filter it can effectively filter out 99 95 solid particles and colloidal substances Size 240 175 2...

Page 9: ...lamp once a year 4 4 ProClean Extra Oral Suction System When not in use store in a dry and ventilated place 4 5 Replacement Filters Items PC2800 SYNFLTR Replacement washable synthetic fiber filter PC...

Page 10: ...on cover pc 2 3 Instruction manual pc 1 7 Disposing notes Following requirements shall be followed as stored Disconnect the appliance from power The components of the product are non toxic dispose of...