Overview
Thank you very much for choosing to the ProClean Extra Oral Suction System. Please read this instruction
manual and the accompanying materials in detail.
The ProClean is covered by a
2 year warranty
(not including replacement filters and uv lamp)
1.1
Contents
This instruction manual includes an overview, instructions for use of the suction unit, operation routines,
maintenance, and precautions. You can view related content as needed and get corresponding help.
1.2
Structure and Composition
The external-oral suction unit is mainly composed of a generally vacuum pressure suction module, a
primary filter, high-efficiency filters, ultraviolet sterilization, plasma sterilization, and a free arm.
1.3
Ideal Applications
It is used for the suction and purification of bacterial aerosols between medical staff and patients during
dental treatment.
1.4
Safety Information
1.4.1
Electrical Safety
●
The power supply for the environment in which this product is used must have a good, standard
grounding device, and the grounding wire must be firm. It must use the national standard single-phase
three-plug power cord, and cooperate with a well-grounded single-phase three-hole socket to ensure good
grounding of this product.
Attention
:
:
:
:
To avoid the risk of electric shock, the device can only be connected to a power cable
with a protective ground. An incorrect grounding plug can cause a risk of electric shock.
●
Before powering on this product, please confirm that the power supply you provide can meet the
input power information marked near the power input port of this product.
●
Do not share the socket with other electrical appliances to prevent the voltage from being unstable
sometimes, which may damage the product
●
When performing maintenance / repair or cleaning this product, be sure to unplug the power cord to
ensure that the product is completely powered off before proceeding.
●
Regularly check the power cord and power plug for damage, and make sure that the power cord is
not crushed by other objects.
1.4.2
Regular Inspection
Regular safety inspections should be performed by trained and skilled personnel, the following items
should be checked at least once a year:
Check the functional status of the equipment and machines.
Check whether the relevant safety signs are clear.
Check whether the performance of the device meets the instructions in the manual.
Test whether the grounding resistance is less than the resistance value 0.1
Ω
required in
GB9706.1
.
1.4.3
Precautions
1)
You should read and understand all contents in the manual carefully before operating.
2)
Follow all warnings and instructions on the instrument during operation.
3)
During maintenance and repair, the water and electricity sources must be emptied before the water and
electricity sources are turned off.
4)
Only trained and skilled medical personnel can operate the equipment. Improper use will cause serious
personal injury.
5)
In following cases, please do not use the device, nor can you open the device or replace the accessories