Name

TPC ProClean Extra Oral Suction 

Model

Model PC2800 

Voltage/Frequency

AC 220V / 50 

Power

1250W 

Noise

70dB 

Suction rate / wind pressure

3000L/min

5KPA

) 

Suction pressure: -10KPa 

Dimensions

368.1*323.5*762MM (not including the free arm) 

Filter level

HEPA 13 

Filter efficiency: 

99.97% 

2.2

 

Classification 

2.2.1

 

Classified by type of anti-motor: Class I equipment 

2.2.2

 

Classification by electric shock protection: ordinary equipment without application part 

2.2.3

 

Classified according to the degree of protection against liquid ingress: IP20 equipment 

2.2.4

 

According to the safety level when used in the case of anesthetic gas mixed with air or flammable 

anesthetic gas mixed with nitrous oxide 

Classification: Not AP type and APG type equipment 

2.2.5

 

Classified by operating mode: intermittent operating equipment 

2.3

 

Transport, storage and use conditions 

2.3.1

 

Transport and storage 

The machine should be transported and stored under following conditions: Temperature: -10 

℃ 

~ + 50 

℃ 

Relative humidity range: 

90%, atmospheric pressure range: 50 kPa ~ 106 kPa 

2.3.2

 

Working condition 

Temperature: + 5 ~ 40 

, relative humidity: 

80%, atmospheric pressure: 86kPa ~ 106kPa 

There should not be a lot of dust and corrosive substances in the surrounding air. 

 

2.4

 

Electromagnetic Compatibility 

Medical electrical equipment requires special tips on electromagnetic compatibility, as well as installation 

and use according to the electromagnetic compatibility information provided in the random files; portable 

and mobile radio frequency communication equipment may affect the use of the equipment. 

In order to ensure the best EMC performance, please use the specified cable. The AC power cord 

specification is 1.8 meters in length; the transducers and cables sold by the company as spare parts for the 

components of the internal active oxygen air disinfection machine. The use of accessories, transducers and 

cables outside the regulations may cause an increase in the emission of the air disinfection machine or 

reduced immunity. 

The suction machine should not be used close to or stacked with other equipment. If it must be used close 

to or stacked, it should be observed to verify that it can operate normally. 

 

Guidance and manufacturer's declaration-electromagnetic emissions 

Dental electric suction machine should be used in the specified electromagnetic environment, the user or 

buyer should ensure that this product is used under such conditions. 

Launch experiment 

Compliance 

Electromagnetic environment-guide 

Radio frequency 

emission 

GB4824 

 

One Unit 

 

Dental electric suction machines use radio frequency 

energy only for their internal functions. Therefore, its radio 

Summary of Contents for PC2800

Page 1: ...ProClean Extra Oral Suction System Model PC2800 TPC 851 Lawson Street City of Industry Ca 91748 P 800 560 8222 www tpcdental com...

Page 2: ...1 Contents 1 Overview 2 2 Technical Specifications 4 3 Product Structure and Installation 6 4 Maintenance 8 5 Trouble shooting 8 6 Electric principle diagram 8 7 Disposing notes 9 8 Packing list 9...

Page 3: ...shock Before powering on this product please confirm that the power supply you provide can meet the input power information marked near the power input port of this product Do not share the socket wit...

Page 4: ...id the use of other highly sensitive instruments and equipment at the same time so as to avoid the equipment from being interfered by electromagnetic sources 1 4 4 Cleaning Please keep the environment...

Page 5: ...air 2 4 Electromagnetic Compatibility Medical electrical equipment requires special tips on electromagnetic compatibility as well as installation and use according to the electromagnetic compatibilit...

Page 6: ...t burst GB T17626 4 2kV On the power cord 1kV For input output lines 2kV On the power cord Not Applicable The grid power supply should have the quality used in a typical commercial or hospital environ...

Page 7: ...joint at the top of the box 3 Install the suction cover onto designated spot as above picture shows 4 Use one phase AC 220V 50hz or other local power 220V 60hz 240V 50hz 110V 60hz would beok 3 3 Oper...

Page 8: ...can effectively remove the droplets and aerosolparticles absorb water vapor Secondary Filter HEPA 13 filter it can effectively filter out 99 95 solid particles and colloidal substances Size 240 175 2...

Page 9: ...lamp once a year 4 4 ProClean Extra Oral Suction System When not in use store in a dry and ventilated place 4 5 Replacement Filters Items PC2800 SYNFLTR Replacement washable synthetic fiber filter PC...

Page 10: ...on cover pc 2 3 Instruction manual pc 1 7 Disposing notes Following requirements shall be followed as stored Disconnect the appliance from power The components of the product are non toxic dispose of...