![Torbal DRX 5 Series Instruction Manual Download Page 8](http://html1.mh-extra.com/html/torbal/drx-5-series/drx-5-series_instruction-manual_1130063008.webp)
Instruction Manual - TORBAL DRX 5 Series
4. Once the pan and scanner have been installed, level the scale by adjusting the rear feet until the level indicator shows
the “air bubble” is in the center position of the sight glass. The level indicator is located on the rear left side of the
scale.
5. After leveling the scale, plug the AC adaptor into the AC adaptor socket located at the rear of the scale.
6. When the AC adaptor is plugged into the wall outlet, the scale will automatically turn on, go through its initialization
process, and calibrate itself. Calibration will initialize automatically whenever the scale is turned on. The scale will
also calibrate whenever the temperature changes by 1 degree Celsius and at two-hour intervals. Calibration will
initialize only if the scale has been stable and idle for 60 seconds, in order not to interrupt weighing operations.
Note: When the AC adapter is plugged into the wall outlet for the first time, the scale will automatically and
frequently initialize temperature calibration as the internal parts of the scale heat up. Once the internal
temperature stabilizes, the scale will initialize the time – interval calibration approximately every two hours. The
scale loses internal temperature only when the unit is completely unplugged from the wall electrical outlet. When
the unit is turned off with the On/Off key, the scale goes into standby mode and internal temperature is
maintained.
8