CHAPTER 19.
DEVIATION WARNING
Deviation Warning:
When enabled the feature allows to check for possible discrepancy between the stored average piece weight, and the
average piece weight of the drug which is counted. The feature signals if a deviation greater than +/- 2% is found.
1. Once command “FILL” is displayed slowly begin to count the pills. In order to check for a possible deviation
you must stop the count between 10 and 15 pieces.
Note: When stopping outside of the 10 to 15 range the feature will not work.
2. When you reach a count of 10 to 15 pieces stop and wait for the scale to stabilize.
3. Once the scale is stable it will calculate a new average piece weight. The new average piece weight is then
compared to the average piece weight which is stored in the scales memory.
4. If a deviation exists which is greater than +/- 2%, but less than +/- 3%, the scale will sound a single beep and “2%
DEV” will appear in the lower left corner of the display. If the deviation is equal to or greater than +/- 3% the
scale will sound three beeps and “3% DEV” will appear in the lower left corner of the display.
58