Instruction Manual - TORBAL DRX 5 Series
CHAPTER 18.
DATA ENTRY REMINDER
When enabled, the feature reminds you to update drug data if a pre-programmed amount of time has passed since
the last entry. Using this feature assures the highest long-term counting accuracy. It is strongly recommended to update
the database periodically as drugs may be reformulated, modified and released under new production lot numbers.
1. When the supply bottle NDC number is scanned, the scale checks the length of time that has passed since the
drug had been last entered or modified in the database. If the length of time is longer than what has been pre-
programmed in the data entry reminder feature (default: 30 days), the scale will display a reminder notice before
proceeding with the counting operation.
2. When the update reminder is displayed, use the navigation keys to toggle between the available options and press
Enter.
Select YES to update the drug information, or NO to postpone or cancel updating the drug data.
Update Average Piece Weight? - YES
1. After selecting
YES
, the scale will allow the drug data to be changed and adjusted.
2. “SAMPLE SIZE 10” will be displayed. To change the sample size, use the left and right navigation keys; to
accept the sample size press
Enter
.
55