79
MAINTENANCE
USER MAINTENANCE ITEM
MANUFACTURER MAINTENANCE ITEMS
Item
Inspection time
Contents
Inspection
Before using
• The instrument works properly.
• The photographing window must be free of
stain or flaw.
Cleaning
When the part is stained
• Photographing window
• External cover, control panel, etc.
Replacement
As required
• Printer form
Item
Checking time
Contents
Cleaning each
component
Within 12 months
• Cleaning outer covers
• Checking the optical system
• Cleaning POWER unit
Operation check
Within 12 months
• Checking the main body operation
• Checking switches
Accuracy check
Within 12 months
• Confirming the corneal endothelium analyzing
functions (using special tools)
• Confirming the cornea thickness measurement
functions (using special tools)