67
SPECIFICATIONS AND PERFORMANCE
SPECIFICATIONS AND PERFORMANCE
SPECIFICATIONS AND PERFORMANCE
Item
Specifications
Corneal endothelium photography
Photography magnification
254x (on the control panel: within ±8%)
Photography range
0.25x0.55mm (within ±8%)
Resolving power
More than 125 line/mm
Fixation target
Central and peripheral
Corneal thickness measurement
Measurement range
0.400-0.750mm
Display unit: 0.001mm step display
Allowance: Less than 0.600mm within ±0.010mm
Over 0.600mm within ±0.025mm
• Corneal thickness is measured only when taking a photograph using
the central fixation target.
NOTE
Essential performance
• Measurement must be performed correctly.
• Monitor screen display must not be distorted.