29
Inspection 4
Inspection Procedure
Inspection Details
Inspection 5
Inspection Procedure
Inspection Details
Is the
inside of the
halogen lamp soiled with
white or black
materials?
The halogen lamp is faulty.
Replace .
28
YES
The halogen lamp is dark.
Wiring check (4)
NO
Is the
output voltage of the
SW power supply normal
(5.1V/15V)?
The SW power supply is
faulty. Replace
.
3
NO
YES
Is the
resistance of the light
adjusting volume normal
(0~30k
Ω
)?
The light adjusting volume is
faulty. Replace
.
30
NO
YES
Is the
external fixation LED
broken?
The external fixation LED is
faulty. Replace
.
26
YES
Wiring check (5)
NO
Is the rectifying
diode broken?
The diode is faulty.
Replace .
25
YES
NO
The external fixation LED
does not light.
R-SLD7-0-0308-11