![Topcon SL-D7 Repair Manual Download Page 23](http://html.mh-extra.com/html/topcon/sl-d7/sl-d7_repair-manual_1128700023.webp)
21
2-3-1 Disassembly and assembly of illumination unit
Illustration
Procedure
1. Removing the slit unit
Disconnect the plug of the lamp house.
Remove the lamp house by turning in the arrow
direction.
Apply the amyl solvent to the top of the screw.
Remove the screws 6S3×16SUS (4 pcs.).
Remove the slit unit by pulling up.
2. Removing the aperture/filter change lever unit
Remove the screws CR2×6SUS (2 pcs.).
Remove the filter handle cover (44770 2601).
* For TAC (44675 2524)
Loosen the screws U2.6×2.5S (2 pcs.).
Plug
Lamp house unit
1
2
Slit unit
Aperture/filter unit
6S3×6SUS
Aperture change gear
CR2×6SUS
Cover
(44770 2601)
(For TAC: 44675 2524)
U2.6×2.5S
Aperture change
(4 pcs.)
(40310 2518)
(2 pcs.)
(2 pcs.)
handle
5
3
1
3
4
1
2
3
4
5
1
2
3
R-SLD7-0-0308-11