12
2-2 Adjustment
* Before carrying out adjustment, if the screw lock has been applied to a screw, dissolve it using Isoamyl acetate.
Illustration
Procedure
1. Focus setting of microscope
Loosen the screws 6S4×12S SUS (3 pcs.) fixing
the microscope on the arm.
Change magnification as looking into the
diopter telescope. Move the microscope back
and forth, tighten the screws 6S4×12S SUS
where it is in focus and apply the screw lock.
Displayed
magnification ×
Diopter
difference (dpt)
25
0.5
16
0.6
10
0.15
(Minimum)
0.35
6S4×12S SUS
(3 pcs.)
Project the image on the focusing rod
and decrease the maximum magnifi-
cation to the minimum.
1
2
R-SLD7-0-0308-11