2-3-4 Trouble shooting (Power supply unit)
Does the
pilot lamp
light
when the power switch
is ON?
2
1
The pilot lamp does not light.
NO
YES
Inspection 1
The pilot lamp blinks.
Inspection 7
Does the
halogen lamp light?
The halogen lamp does not light.
NO
YES
Inspection 2
Does the
light intensity of
the halogen lamp change
continuously by using the
light adjusting
volume
?
30
The light intensity of the halogen
lamp does not change continuously.
NO
YES
Inspection 3
Is the
brightness of the halogen
lamp normal?
The halogen lamp is dark.
NO
YES
Inspection 4
Does the external
fixation lamp light?
The external fixation LED does not
light.
NO
YES
Inspection 5
Is R/L detected?
R/L is not detected.
NO
YES
Inspection 6
Normal