9. Slit image shift setting when rotating the slit
illumination unit horizontally (in 30° right and
left each)
Put the eyepiece with scale into the microscope
unit and adjust the diopter with the diopter
adjusting ring.
Project the aperture of
φ
0.2 on the focusing rod.
Watch the deviation as moving the illumination
arm in 30° right and left and, after loosening the
screws V3×8S (2 pcs.) of the mirror bracket,
adjust to keep the right and left movement
within 0.1.
After adjustment, tighten the screws fully and
apply the screw lock.
10. Slit image shift setting when inclined
Put the eyepiece with scale into the microscope
unit and adjust the diopter with the diopter
adjusting ring.
Project the aperture of
φ
0.2 on the focusing rod.
Watch the amount of movement as moving the
inclination unit as shown by the arrow.
Make adjustment with the screws V3×8S (2
pcs.) and U3×3S (2 pcs.) of the mirror bracket.
φ
0.2
V3×8S (2 pcs.) Microscope unit
Eyepiece with
30°
Illumination arm
(Right eye)
[Movement: Within 0.1mm]
V3×8S (2 pcs.)
U3×3S (2 pcs.)
(Right eye)
[Movement: Within 0.05]
scale
30°
3
1
3
2
1
2
3
4
1
2
3
4
R-SLD7-0-0308-11
Summary of Contents for SL-D7
Page 1: ...ISSUED AUG 2003 REVISED REPAIR MANUAL SL D7 SLIT LAMP TOPCON CORPORATION R SLD7 0 0308 11 ...
Page 5: ...3 1 3 Nomenclature For Japan R SLD7 0 0308 11 ...
Page 7: ...5 For TMS TE TS and General R SLD7 0 0308 11 ...
Page 35: ...33 2 Halogen lamp cord R SLD7 0 0308 11 ...
Page 36: ...34 3 Halogen lamp and socket R SLD7 0 0308 11 ...
Page 37: ...35 4 External fixation target 5 Cross slide unit CN1 CN2 R SLD7 0 0308 11 ...