4. Communication to PC
4.1
Communication command
The RD-80SA can be communicated with PC via RS-232C and LAN interface.
User commands for creating program by the customer are listed below.
4.2
Command list
Communication commands are as follows;
Communication command
Function
RM
Enters to the remote mode
LM
Enters to the local mode.
(Returns from the remote mode.)
CAL
Conducts calibration
WHO
Reads Model name
VER
Reads Software version
SRL
Reads serial number
ST
Starts measurement.
After completing measurement, measured data are sent
from the instrument.
STC
Starts continuous measurement. Each measured data is
not sent from the instrument.
SPC
Stops continuous measurement
SMS
Reads the status of continuous measurement
0:Normal stop 1:Continuous measurement
3:Abnormal stop
SMD
Reads latest measured data in continuous measurement
MV_#
Sets On or Off of average measurement.
0
:
SINGLE
1
:
AVERAGE
AV_#
Sets measurement times for average measurement
Valid value range
:
2 - 20
MG_#
Selects measurement range mode
0
:
AUTO FULL
1
:
AUTO FIRST
2
:
AUTO ADJUST
3
:
RANGE ALL
4
:
RANGE EACH
SR_#
Sets common measurement range in [RANGE ALL]
Valid value range
:
1 - 8
FO_#
Sets measurement range for OPEN filter in [RANGE
EACH]
Valid value range
:
1 - 8