TH2523 Series Operation Manual
Ver1.2
57
Return:
50 the supply frequency is 50Hz
60 the supply frequency is 60Hz
7.1.3
APERture Subsystem Commands
The APERture subsystem commands are mainly used to set the measurement speed and the
average times used in measurement.
The APERture? query returns the current measurement speed and average times used in
measurement.
Command Syntax:
FAST
APERture
MEDium
[,<value>]
SLOW
Where:
FAST
:
Fast: 100meas/sec
MEDium
:
Medium: 50meas/sec
SLOW
:
Slow: 6.25meas/sec
<value>
Average number ranging from 1 to 128(NR1).
For example: WrtCmd(“APER MED,55”);
Query syntax: APERture?
Return format:
FAST
MED
,<NR1><NL^END>
SLOW
7.1.4
TRIGger Subsystem Commands:
The TRIGger subsystem commands set the instrument trigger source and the delay time after
triggering, and triggers instrument measurement.
Command Tree: