8
9
EN
EN
SLIT LAMP TSL-7000Z by TOMEY
SLIT LAMP TSL-7000Z by TOMEY
2. Using the wrench provided, fi t the headrest assembly to the tabletop using the hex bolts
and washers. The headrest assembly locates on the underside of the tabletop. Take care
not to over tighten the hex bolts.
3. Attach the patient grab handles (7) to the headrest assembly.
4. Place the Slit Lamp base on the runners. Ensure that the wheels are in line with each
other. Check that the guide wheels are tight.
5. Fit the runner covers to the runners by gently sliding them inwards, towards each other.
5.2 ILLUMINATION TOWER ASSEMBLY PROCEDURE
1. Remove the hex bolt (a) from the base of the Illumination Tower, and then place the
Illumination Tower on the Slit Lamp base with the base notch (b) and pin (c) aligned.
Attach the tower to the base using the hex bolt removed earlier and tighten using the
wrench provided.
2. Carefully fi t the microscope body to the arm – ensuring it is pushed to the stop. Tighten
using the securing knob on the side.
a
b
c
slide to stop and secure
31. Main power switch
32. Power supply unit
33. Glide plate
5. ASSEMBLY
Your Tomey Slit Lamp has been designed to fi t on to an electric insulated medical table base or on
to an electric insulated and fi re resistant medical table top, e.g. a refraction stand or combi unit.
Take care when unpacking your Slit Lamp that you do not accidentally damage or discard any
of the contents.
Leave the Slit Lamp in the packing for several hours after delivery before
unpacking to reduce the risk of condensation forming.
Tomey Slit Lamps can be fi tted to most Refraction Stands / Combi units. Tomey advises that this
shall be carried out by a suitably trained technicians to ensure performance and safety are not
affected.
The Refraction Stand, Combi Unit or table leg must be IEC 60601-1
compliant.
If you are fi tting or have fi tted your Slit Lamp to a medical or Tomey table leg / base ensure it is
situated on a fi rm and level fl oor.
If the table leg / base has castors ensure the following before moving it to another location:
1. The table is at its lowest position.
2. The power cord is removed.
3. The Slit Lamp arm and base locking knobs are tightened.
4. The runner covers are securely located.
5. The system is moved by grasping it at its lowest practical point.
5.1 TABLE TOP AND BASE ASSEMBLY PROCEDURE
1. Attach the Slit Lamp table top to your table leg using the M6 x 20mm screws and
washers. Note that the power supply and accessory drawer should face the operator.
The security of the fi tment of the table top to the table leg is critical for
patient and Slit Lamp safety
32
33
31
7
Runners