SLIT LAMP TSL-7000Z by TOMEY

SLIT LAMP TSL-7000Z by TOMEY

to the same or a similar device with a visible light source within 24 hours. 

This applies, in particular, if the retina has been photographed with a 

fl ashbulb in advance. 
Tomey GmbH shall on request, provide the user with a graph showing the 

relative spectral output of the instrument.

2.2  WARNINGS AND CAUTIONS

Please note that the proper and safe functioning of our instruments is only guaranteed if both 

the instruments and their accessories are exclusively from Tomey GmbH. The use of other 

accessories may result in increased electromagnetic emissions or reduced electromagnetic 

immunity of the device and may lead to incorrect operation.
Observe the following prescriptions in order to ensure safe operation of the instrument.

WARNINGS

•  Never use the instrument if visibly damaged and periodically inspect it for signs of damage 

or misuse.

•  Check your Tomey product for signs of transport / storage damage prior to use.
•  Do not use in the presence of fl ammable gases / liquids, or in an oxygen rich environment.
•  US Federal Law restricts this device to sale by or on the order of a physician or practitioner.
•  This device is intended to be used only by suitably trained and authorised healthcare 

professionals.

•  This product should not be immersed in fl uid.
•  Repairs and modifi cations to the instrument must be made only by the specialized 

technicians of the manufacturer’s Technical Service Centre or by personnel trained and 

authorised by the manufacturer. The manufacturer declines any and all responsibility for 

loss and/or damages resulting from unauthorised repairs; furthermore, any such actions 

will invalidate the warranty.

•  The power switch and mains plug are the means of isolating the device from the mains 

supply - ensure both the power switch and mains plug are accessible at all times.

•  Do not position the equipment so that is diffi cult to press the power switch or remove the 

mains plug from the wall socket.

•  Refraction stand variants or adaptors should only be used in combination with EN/IEC 

60601-1 and EN/IEC 60601-1-2 compliant power supplies and devices. 

•  Route power cords safely to eliminate risk of tripping or damage to user.

•  Before any cleaning of the instrument or the base unit ensure the power lead is 

disconnected.

•  LEDs can reach high temperatures in use – allow to cool before handling.

•  Do not exceed maximum recommended exposure time.

•  Should the instrument suffer shocks (for example, should it accidentally fall), and the 

optical system or the illumination system are damaged it may be necessary to return the 

instrument to the manufacturer for repair.

•  After removal of the LED, do not touch the Slit Lamp LED electrical contacts and patient 

simultaneously.

•  The owner of the instrument is responsible for training personnel in its correct use.
•  Ensure the instrument or instrument table is placed on a level and stable surface.
•  Use only genuine Tomey approved parts and accessories or device safety and performance 

may be compromised.

•  Shut down after every use. In case the dust cover is used: risk of overheating.
•  For indoor use only (protect from moisture).
•  Electrical equipment can be affected by electromagnetic interference. If this occurs whilst 

using this equipment, switch the unit off and reposition.

•  Do not touch accessible connectors and the patient simultaneously.
•  The Tomey Slit Lamp should not be used adjacent to or stacked with other equipment. If 

adjacent or stacked use with other equipment is necessary the Tomey Slit Lamp should be 

observed/monitored to verify normal operation in the confi guration in which it will be used.

•  The Tomey Slit Lamp cannot be used in the vicinity of sources known to cause 

electromagnetic disturbance (magnetic resonance imaging, computed tomography, 

radio-frequency identifi cation, metal detectors, electronic article surveillance and other 

electromagnetic security systems). Do not bring the Tomey digital Slit lamp into magnetic 

resonance environment.

Before use, the Slit Lamp should be allowed to adjust to the ambient 

room temperature for several hours. This is especially important when the 

unit has been stored or transported in a cold environment; this can cause 

severe condensation to develop on the optical elements.

2.3  CONTRAINDICATION

There is no restriction to patient population this device can be used with other than those 

outlined in the contra-indications stated below. 
Slit lamps can produce discomfort in some photophobic patients due to the high illumination. 

Additionally patients must be co-operative and capable of sitting upright for the duration of 

the examination, therefore the technique may not be suitable for patients who are unable to sit 

upright for long periods of time or those with limited neck and back mobility.

Summary of Contents for TSL-7000Z

Page 1: ...TSL 7000Z SLIT LAMP OPERATOR MANUAL Z Series...

Page 2: ...IONS FOR USE 12 6 1 SETTING THE BINOCULARS 12 6 2 PREPARING THE PATIENT AND USING THE SLIT LAMP 13 6 3 DESCRIPTION OF FILTERS APERTURES AND MAGNIFICATIONS 14 7 ROUTINE MAINTENANCE 15 7 1 LED SYSTEMS 1...

Page 3: ...f the anterior eye segment The Tomey Slit Lamp is an instrument consisting of a light source that can be focused to shine a thin sheet slit of light into the eye It is used in conjunction with a biomi...

Page 4: ...is disconnected LEDs can reach high temperatures in use allow to cool before handling Do not exceed maximum recommended exposure time Should the instrument suffer shocks for example should it accident...

Page 5: ...f used cleaning materials 4 NAMES OF CONTROLS AND COMPONENTS Headrest Assembly 1 Fixation light cable 2 Forehead rest band 3 Patient s eye height marker 4 Fixation light 5 Chinrest 6 Chinrest height a...

Page 6: ...an electric insulated and fire resistant medical table top e g a refraction stand or combi unit Take care when unpacking your Slit Lamp that you do not accidentally damage or discard any of the conten...

Page 7: ...ETERS T TYPE AND R TYPES Tomey Applanation Tonometer T TYPE 1 Position the guide plate in the tonometer test bar support hole on the slit lamp 2 Lift Tonometer out of the packaging and assemble it by...

Page 8: ...tact with the patient should be cleaned in accordance with these instructions prior to the examination Tomey recommends the use of disposable hygienic chinrest tissues on the Chinrest before the patie...

Page 9: ...3 In the unlikely event of an LED failure please contact Tomey or your local distributor for guidance on the replacement procedure 7 2 REGULARLY INSPECT THE DEVICE FOR DAMAGE OR DIRT 1 Routinely clean...

Page 10: ...ded for use in home environment Harmonic emissions IEC 61000 3 2 Class B Voltage fluctuations flicker emissions IEC 61000 3 3 Complies 9 2 ELECTROMAGNETIC IMMUNITY Guidance and manufacturer s declarat...

Page 11: ...150kHz to 80 MHz field strengths should be less than 10 V m 9 3 RECOMMENDED SAFE DISTANCES Recommended separation distances between portable and mobile RF communications equipment and the Tomey Slit L...

Page 12: ...al Electrical System as defined in EN IEC 60601 1 2006 The power supply forms a part of ME equipment Fuse Ratings and Quantity 2 5 amp anti surge Fuse current 2 5A Voltage rating V AC 250V Breaking ca...

Page 13: ...k Patient Handle EP79 80070 11 PACKAGING AND DISPOSAL INFORMATION Applicable in the European Union and other European Countries with separate Collection Systems Disposal of old electrical and electron...

Page 14: ...AA UK Freephone 0800 521251 Tel 44 0 1753 857177 Fax 44 0 1753 827145 EP59 51500 Issue 5 Date of Issue 12 05 2021 SUPPLIER TOMEY GmbH Wiesbadener Stra e 21 90427 N rnberg Germany Tel 49 911 938 546 2...

Reviews: