SLIT LAMP TSL-7000Z by TOMEY

SLIT LAMP TSL-7000Z by TOMEY

Consult instructions for use

General warning sign

Date of manufacture

Warning: Electricity

Manufacturer’s name and 

address

Warning: Floor level obstacle

Country of manufacture

Warning: Non-ionizing 

radiation

Waste Electrical and Electronic 

Equipment (WEEE) recycling

Warning: Optical radiation

This way up

Warning: Hot surface

Keep dry

Conformité Européene

Fragile

Type B applied part

Do not use if package is 

damaged

Class II equipment

Temperature limit

Atmospheric pressure 

limitation

Authorised representative in 

the European Community

Humidity limitation

Catalogue number

Serial number

Translation

Medical device

The Tomey Slit Lamp is designed and built in conformity with Directive 93/42/EEC, Regulation (EU) 2017/745 and 

ISO 13485 Medical Devices Quality Management Systems.
Classifi cation:  CE:  Class I
 

FDA: Class II

The information contained within this manual must not be reproduced in whole or part without the manufacturer’s 

prior written approval. As part of our policy for continued product development we the manufacturer reserve the right 

to make changes to specifi cations and other information contained in this document without prior notice.
Copyright © Tomey GmbH 2021. Published in the UK 2021.

1.  INDICATIONS FOR USE 

This device is intended to be used only by suitably trained and authorised healthcare 

professionals.

CAUTION: Federal Law restricts this device to sale by or on the order of a 

physician or practitioner.

Intended use / purpose of instrument

This Tomey Slit Lamp is an AC-powered slit lamp biomicroscope and is intended for use in eye 

examination of the anterior eye segment, from the cornea epithelium to the posterior capsule. 

It is used to aid in the diagnosis of diseases or trauma which affects the structural properties of 

the anterior eye segment.
The Tomey Slit Lamp is an instrument consisting of a light source that can be focused to shine 

a thin sheet (slit) of light into the eye. It is used in conjunction with a biomicroscope. The lamp 

facilitates an examination of the anterior segment, or frontal structures and posterior segment,
of the human eye, which includes the eyelid, sclera, conjunctiva, iris, natural crystalline lens, and 

cornea. The binocular Slit Lamp examination provides stereoscopic magnifi ed view of the eye 

structures in detail, enabling anatomical diagnoses to be made for a variety of eye conditions.

Brief description of the instrument

This Tomey Slit Lamp can either be mounted onto a custom table top supplied by Tomey or can 

be mounted on a third parties table top (refraction unit) by suitably trained technicians.
The Tomey Slit Lamp consists of 5 assemblies; Illumination Tower; Observation System; XYZ 

Translation Base; Chinrest Assembly and a Table Top with power supply and accessory drawer.
The light intensity is controlled by a variable rheostat located on the XYZ Translation Base. 

There are a number of selectable fi lters allowing the user to control the characteristics of the 

examination light. Many ophthalmic instruments illuminate the eye with intense light. The light 

intensity on the Tomey Slit Lamp is continuously adjustable from maximum to zero.

2.  SAFETY

2.1  PHOTOTOXICITY

CAUTION: The light emitted from this instrument is potentially hazardous. 

The longer the duration of exposure, the greater the risk of ocular 

damage. Exposure to light from this instrument when operated at 

maximum intensity will exceed the safety guideline after 81 seconds.
While no acute optical radiation hazards have been identifi ed for slit 

lamps, we recommend keeping the intensity of the light reaching the 

patient’s retina to the minimum possible for the respective diagnosis. 

Children, people with aphakia and people suffering from eye conditions 

are most at risk. An increased risk may also occur if the retina is exposed 

Summary of Contents for TSL-7000Z

Page 1: ...TSL 7000Z SLIT LAMP OPERATOR MANUAL Z Series...

Page 2: ...IONS FOR USE 12 6 1 SETTING THE BINOCULARS 12 6 2 PREPARING THE PATIENT AND USING THE SLIT LAMP 13 6 3 DESCRIPTION OF FILTERS APERTURES AND MAGNIFICATIONS 14 7 ROUTINE MAINTENANCE 15 7 1 LED SYSTEMS 1...

Page 3: ...f the anterior eye segment The Tomey Slit Lamp is an instrument consisting of a light source that can be focused to shine a thin sheet slit of light into the eye It is used in conjunction with a biomi...

Page 4: ...is disconnected LEDs can reach high temperatures in use allow to cool before handling Do not exceed maximum recommended exposure time Should the instrument suffer shocks for example should it accident...

Page 5: ...f used cleaning materials 4 NAMES OF CONTROLS AND COMPONENTS Headrest Assembly 1 Fixation light cable 2 Forehead rest band 3 Patient s eye height marker 4 Fixation light 5 Chinrest 6 Chinrest height a...

Page 6: ...an electric insulated and fire resistant medical table top e g a refraction stand or combi unit Take care when unpacking your Slit Lamp that you do not accidentally damage or discard any of the conten...

Page 7: ...ETERS T TYPE AND R TYPES Tomey Applanation Tonometer T TYPE 1 Position the guide plate in the tonometer test bar support hole on the slit lamp 2 Lift Tonometer out of the packaging and assemble it by...

Page 8: ...tact with the patient should be cleaned in accordance with these instructions prior to the examination Tomey recommends the use of disposable hygienic chinrest tissues on the Chinrest before the patie...

Page 9: ...3 In the unlikely event of an LED failure please contact Tomey or your local distributor for guidance on the replacement procedure 7 2 REGULARLY INSPECT THE DEVICE FOR DAMAGE OR DIRT 1 Routinely clean...

Page 10: ...ded for use in home environment Harmonic emissions IEC 61000 3 2 Class B Voltage fluctuations flicker emissions IEC 61000 3 3 Complies 9 2 ELECTROMAGNETIC IMMUNITY Guidance and manufacturer s declarat...

Page 11: ...150kHz to 80 MHz field strengths should be less than 10 V m 9 3 RECOMMENDED SAFE DISTANCES Recommended separation distances between portable and mobile RF communications equipment and the Tomey Slit L...

Page 12: ...al Electrical System as defined in EN IEC 60601 1 2006 The power supply forms a part of ME equipment Fuse Ratings and Quantity 2 5 amp anti surge Fuse current 2 5A Voltage rating V AC 250V Breaking ca...

Page 13: ...k Patient Handle EP79 80070 11 PACKAGING AND DISPOSAL INFORMATION Applicable in the European Union and other European Countries with separate Collection Systems Disposal of old electrical and electron...

Page 14: ...AA UK Freephone 0800 521251 Tel 44 0 1753 857177 Fax 44 0 1753 827145 EP59 51500 Issue 5 Date of Issue 12 05 2021 SUPPLIER TOMEY GmbH Wiesbadener Stra e 21 90427 N rnberg Germany Tel 49 911 938 546 2...