SLIT LAMP TSL-7000Z by TOMEY

SLIT LAMP TSL-7000Z by TOMEY

Table top dimensions

405 x 500mm

Fixation lamp

LED

Light source

LED

LED output power

 240Klux (+/-20%)

Weight, packed  (approx.)

Slit Lamp complete

25Kg, 90 x 58 x 45cm W x D x H

Protection against ingress 

IPxO

Class II ME equipment

Insulation between mains parts and the functional earth provide at least two means of protection.

Power Supply 

Power supply unit

Switch mode, (100V-240V input) +/- 10% multi plug compliant to EN/

IEC 60601-1 EN/IEC 61000-6-2, EN/IEC 61000-6-3

Fuse

T2.5AH, 250V

Power supply output

12V DC: 2.5 amps must be EN/IEC 60601 compliant

Complies with

Electrical Safety (Medical) EN/IEC 60601-1
Electromagnetic compatibility EN/IEC 60601-1-2
Ophthalmic instruments - Fundamental requirements and test methods 

ISO 15004-1
Ophthalmic instruments - Optical radiation hazard ISO 15004-2

When the Slit Lamp is connected to the power supply together the constitute a Medical 

Electrical System as defined in EN/IEC 60601-1:2006.
The power supply forms a part of ME equipment.

Fuse Ratings and Quantity

2.5 amp anti-surge  

Fuse current 2.5A  

Voltage rating V AC 250V 

Breaking capacity 1500Amps  

Blow characteristic: Time Delay

Environmental Conditions:

USE

35

°C

10

°C

90%

30%

1060 hPa

800 hPa

Shock (without packing)

10 g, duration 6 ms

STORAGE CONDITIONS

*50

°C

-10

°C

95%

10%

1060 hPa

700 hPa

TRANSPORT CONDITIONS

*50

°C

-40

°C

95%

10%

1060 hPa

500 hPa

Vibration, sinusoidal

10 Hz to 500 Hz: 0.5g

Shock

30 g, duration 6 ms

Bump

10 g, duration 6 ms

*This instrument does not meet the temperature requirements of ISO 15004-1 for storage and transportation. Do 

not store or transport this instrument in conditions where the temperature may rise above 50 °C.

10. ACCESSORIES AND SPARES

Item

Part Number

DSL Digital Background Illum.Assy

3020-P-5039

Tomey PSU 12V

3020-P-5040-TOM

DSL Accessory Kit Packaging

3020-P-5221

DSL Screw Kit

3020-P-5240

Refraction Cable (XY-Bulb)

3020-P-7013

DSL Refraction Stand Cable Kit

3020-P-7014

DSL Digital Backlight Filters

3020-P-7054

1m DSL to PSU Cable

3020-P-7068

Tomey user instruction USB

3020-P-7069

Summary of Contents for TSL-7000Z

Page 1: ...TSL 7000Z SLIT LAMP OPERATOR MANUAL Z Series...

Page 2: ...IONS FOR USE 12 6 1 SETTING THE BINOCULARS 12 6 2 PREPARING THE PATIENT AND USING THE SLIT LAMP 13 6 3 DESCRIPTION OF FILTERS APERTURES AND MAGNIFICATIONS 14 7 ROUTINE MAINTENANCE 15 7 1 LED SYSTEMS 1...

Page 3: ...f the anterior eye segment The Tomey Slit Lamp is an instrument consisting of a light source that can be focused to shine a thin sheet slit of light into the eye It is used in conjunction with a biomi...

Page 4: ...is disconnected LEDs can reach high temperatures in use allow to cool before handling Do not exceed maximum recommended exposure time Should the instrument suffer shocks for example should it accident...

Page 5: ...f used cleaning materials 4 NAMES OF CONTROLS AND COMPONENTS Headrest Assembly 1 Fixation light cable 2 Forehead rest band 3 Patient s eye height marker 4 Fixation light 5 Chinrest 6 Chinrest height a...

Page 6: ...an electric insulated and fire resistant medical table top e g a refraction stand or combi unit Take care when unpacking your Slit Lamp that you do not accidentally damage or discard any of the conten...

Page 7: ...ETERS T TYPE AND R TYPES Tomey Applanation Tonometer T TYPE 1 Position the guide plate in the tonometer test bar support hole on the slit lamp 2 Lift Tonometer out of the packaging and assemble it by...

Page 8: ...tact with the patient should be cleaned in accordance with these instructions prior to the examination Tomey recommends the use of disposable hygienic chinrest tissues on the Chinrest before the patie...

Page 9: ...3 In the unlikely event of an LED failure please contact Tomey or your local distributor for guidance on the replacement procedure 7 2 REGULARLY INSPECT THE DEVICE FOR DAMAGE OR DIRT 1 Routinely clean...

Page 10: ...ded for use in home environment Harmonic emissions IEC 61000 3 2 Class B Voltage fluctuations flicker emissions IEC 61000 3 3 Complies 9 2 ELECTROMAGNETIC IMMUNITY Guidance and manufacturer s declarat...

Page 11: ...150kHz to 80 MHz field strengths should be less than 10 V m 9 3 RECOMMENDED SAFE DISTANCES Recommended separation distances between portable and mobile RF communications equipment and the Tomey Slit L...

Page 12: ...al Electrical System as defined in EN IEC 60601 1 2006 The power supply forms a part of ME equipment Fuse Ratings and Quantity 2 5 amp anti surge Fuse current 2 5A Voltage rating V AC 250V Breaking ca...

Page 13: ...k Patient Handle EP79 80070 11 PACKAGING AND DISPOSAL INFORMATION Applicable in the European Union and other European Countries with separate Collection Systems Disposal of old electrical and electron...

Page 14: ...AA UK Freephone 0800 521251 Tel 44 0 1753 857177 Fax 44 0 1753 827145 EP59 51500 Issue 5 Date of Issue 12 05 2021 SUPPLIER TOMEY GmbH Wiesbadener Stra e 21 90427 N rnberg Germany Tel 49 911 938 546 2...

Reviews: