8-8
Table 3
【
Input a.c. power PORT
】
Table 4
【
Signal input / output parts PORT
】
Test
Standard
IMMUNITY TEST LEVELS
HOME HEALTHCARE ENVIRONMENT
Electrical fast
transients / bursts
IEC 61000-4-4
土
2kV
100kHz repetition frequency
Surges
Line-to-line
IEC 61000-4-5
土
0.5 kV
,土
1 kV
Surges
Line-to-ground
IEC 61000-4-5
土
0.5 kV
,土
1kV
,土
2kV
Conducted
disturbances induced
by RF fields
IEC 61000-4-6
3V
0.15MHz - 80MHz
6V in ISM and amateur radio bands between
0.15MHz and 80MHz
80% AM at 1kHz
Voltage dips
IEC 61000-4-11 0%
UT
; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270° and
315°
0%
UT
; 1 cycle
and
70%
UT
; 25 / 30 cycles
Single phase: at 0°
Voltage interruptions IEC 61000-4-11 0%
UT
; 250 / 300 cycle
Test
Standard
IMMUNITY TEST LEVELS
HOME HEALTHCARE ENVIRONMENT
ELECTROSTATIC
DISCHARGE
IEC 61000-4-2
土
8 kV contact
土
2 kV
,土
4 kV
,
土
8 kV
,
土
15 kV air
Electrical fast
transients /
bursts
IEC 61000-4-4
土
1 kV
100 kHz repetition frequency
Conducted
disturbances induced
by RF fields
IEC 61000-4-6
3 V
0.15 MHz - 80 MHz
6 V in ISM and amateur radio bands between
0.15 MHz and 80 MHz
80 % AM at 1 kHz
Summary of Contents for MR-6000
Page 2: ......
Page 24: ... 1 18 This page is intentionally left blank ...
Page 30: ... 2 6 This page is intentionally left blank ...
Page 46: ... 4 2 This page is intentionally left blank ...
Page 56: ... 6 4 This page is intentionally left blank ...
Page 58: ... 7 2 This page is intentionally left blank ...
Page 67: ...8 9 This page is intentionally left blank ...
Page 68: ......
Page 70: ...0120 ...