11
6. Various functions
1. Auto Zero function (torque)
In RUN mode, it will be automatically adjusted to zero by pushing Clear key.
(Auto Zero function is applicable only when the torque load is within 7.5% of the
max range). If the loaded torque is over 7.5%, the display shows “Err9”.
In case of Err9, refer to
③
Error message.
2. Angle speed check / zero adjustment function
When turning on the power, angle speed check will be automatically conducted to
make zero adjustment. During zero adjustment, CTA2-G should be put still (do not
move) otherwise the display may show “Err0”. The angle speed check will also be
automatically processed when it is in still kept still for more than 2 seconds.
In case of Err0, refer to
③
Error message.
During angle speed check, the display shows” “.
OK
Error
3. Error message
Error messages will appear on the display when there is some error during 1 or
2 process.
<< Err 0 >>
Error in angle speed detection.
◎
Turn off the power once. Keep the body still (no movement) and turn on the
power again.
- If Err disappears, then it operates normally.
- If Err message remains on the display, contact TOHNICHI or the nearest
distributor to ask for repair.