Operating Instruction
TNI
soft
Flow
50
homecare system
Recommended separation distances between
portable and mobile RF communications equipment and the “
TNI
soft
Flow
50
The
TNI
soft
Flow
50
is intended for use in the electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the “
TNI
soft
Flow
50
” can help
prevent electromagnetic interference by maintaining a miminum distance between portable
and mobile RF communications equipment (transmitters) and the “
TNI
soft
Flow
50
” as recom-
mended below, according to the maximum output power of the communication equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d = 1.2√P
80 MHz to 800 MHz
d = 1.17√P
800 MHz to 2.5
GHz
d = 2.3√P
0.01
0.12
0.12
0.23
0.1
0.38
0.37
0.73
1
1.2
1.2
2.3
10
3.8
3.7
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separa-
tion distance d in metres (m) can be determined using the equation applicable to the frequency
of the transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is af-
fected by absorption and reflection from structures, objects and people.
Table 9 - Recommended separation distances between
portable and mobile RF communications equipment and the “
TNI
soft
Flow
50
”