12) Electromagnetic compatibility (EMC)
SAFETY NOTE
•
The
TNI
soft
Flow
50
is a MEDICAL ELECTRICAL EQUIPMENT and needs special precautions
regarding EMC and needs to be installed and put into service according to the EMC informa-
tion provided in the ACCOMPANYING DOCUMENTS.
•
Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL EQUIP-
MENT!
•
The
TNI
soft
Flow
50
should not be used adjacent to or stacked with other equipment. If
adjacent or stacked use is necessary the
TNI
soft
Flow
50
should be observed to verify nor-
mal operation in the configuration in which it will be used.
•
The use of other accessories , cables or converters to the device
TNI
soft
Flow
50
can increase
the emission and reduce the immunity of the
TNI
soft
Flow
50
.
•
In accordance to the applicable standard the
TNI
soft
Flow
50
has no essential performance.
•
The
TNI
soft
Flow
50
may be interfered with by other equipment, even if that other equip-
ment complies with CISPR EMISSION requirements.
Guidance and manufacturer’s declaration - electromagnetic emissions
The
TNI
soft
Flow
50
is intended for use in the electromagnetic environment specified below.
The customer or the user of the
TNI
soft
Flow
50
should assure that it is used in such an
environment.
Emissions test
Compliance
Electromagnetic environment - guid-
ance
RF emissions
CISPR 11
Group 1
The
TNI
soft
Flow
50
uses RF energy
only for its internal function. Therefore,
its RF emissions are very low and are not
likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The
TNI
soft
Flow
50
is suitable for use
in all establishments, including domestic
establishments and those directly con-
nected to the public low-voltage power
supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Complies
Table 6 -Guidance and manufacturer’s declaration - electromagnetic emissions