Operating Instruction
TNI
soft
Flow
50
homecare system
REAL VALUES on the display
•
The first line shows the humidity (dew point) real value
•
and
↑
o
the system heats up,
•
and
↓
o
the system cools down.
•
The second line shows the Flow rate real value.
o
this value corresponds to the current exact, regulated flow rate released in l/min,
o
consisting of air or air-oxygen mixture.
•
Additionally, the admixture of oxygen:
o
O
2
admixture in l/min,
o
and the resulting oxygen (concentration) in %, FiO
2
-value.
4.1.2) Setting humidity (dew point)
WARNING
Humidity and dew point
•
In order to set a nominal value, the system requires a setup-time of about <30 minutes.
•
If the humidity value is altered (no matter if increasing or decreasing the value), the system
will always require a physical setup-time of about 10 minutes.
•
The system is to be used in the recommended environment (see chapter “Technical data”).
SAFETY NOTE
•
The Real value is obligatory for therapy.
•
If your nose feels dry during therapy,
o
check for correct choice and setting of the respective humidifier type,
o
make sure that the “humidifier homecare complete” contains enough water,
o
increase humidity.
•
If you feel a slight stinging in the nose,
o
check for the correct choice and setting of the respective humidifier type,
o
make sure that the “humidifier homecare complete” contains enough water,
o
increase humidity.
•
If condensate forms in the applicator tube or interfering water droplets spurt out of the
applicator,
o
check for correct choice and setting of the respective humidifier type,
o
check for correct ambient conditions. The room temperature might be too low,
o
reduce humidity; humidity might be set too high for the valid ambient conditions.